PN B38322AA
6-37
Functional Pages
Max Temperature Offset Page
6
Max Temperature Offset Page
Figure 6.39
Max Temperature Offset Page
Select the
Max Temperature Offset
button on the Basic tab of the
to display the
Max Temperature Offset
page. Use this page to set the tolerable deviation above the set temperature
during a run. Once the tolerable deviation above the set temperature has been selected, the user can
specify whether the run should shut down or continue if the overtemp setting is reached by using
the
Max Temperature Shutdown
button on the Basic tab of the
.
In addition to the Header Bar and the Footer Bar, the
Max Temperature Offset
page contains the
following elements:
•
The
Tolerable Deviation Above the Set Temperature
page shows the current setting in degrees
Celsius. The system default is 4°C.
•
Use the keypad to change the setting. You can use the
Clear
and
Back
keys to make corrections.
•
The
Cancel
button discards your changes and dismisses the page.
•
The
OK
button accepts your changes and dismisses the page.
Summary of Contents for Avanti JXN-26
Page 4: ...PN B38322AA iv Revision History...
Page 30: ...PN B38322AA xxx Introduction Symbols Symbols...
Page 32: ...PN B38322AA xxxii Introduction Graphics...
Page 44: ...PN B38322AA 1 12 Description Available Rotors...
Page 54: ...PN B38322AA 2 10 The Touch Screen Interface About Your Instrument...
Page 154: ...PN B38322AA 6 60 Functional Pages Zonal Operation Page...
Page 168: ...PN B38322AA 7 14 Maintenance and Troubleshooting Supply List...
Page 176: ...PN B38322AA A 8 Preinstallation Requirements Using J2 Series Rotors in the Avanti JXN Series...
Page 194: ...PN B38322AA D 12 Third Party Contributions and Legal Notices Contributions and Legal Notices...
Page 198: ...PN B38322AA Abbreviations 2 Abbreviations...
Page 206: ...Index 8 Index...
Page 208: ...PN B38322AA Warranty 2 Beckman Coulter Inc Avanti JXN Series Centrifuge Warranty...
Page 209: ......