PN B10093AE
A-7
Preinstallation Requirements
Using J2 Series Rotors in the Avanti J-26S XPI
A
Using J2 Series Rotors in the Avanti J-26S XPI
CAUTION
Do not use the Beckman Coulter JA-10, JS-7.5, JA-14, or JS-13 rotors in the
Avanti J-26S XPI centrifuge before reading this information.
Checking for Rotor Drive Pins
Rotors used in Avanti J series centrifuges must have drive pins in the rotor drive hole (see
). These drive pins engage with the centrifuge spindle hub to ensure that the rotor does
not slip during acceleration. Some Beckman Coulter rotors, including the JA-10, the JS-7.5, the
JA-14, and the JS-13, have been manufactured without drive pins because pins were not needed
when these rotors were used in J2 series centrifuges.
Figure A.5
Checking the Rotor for Drive Pins
1.
Top View
2.
Side View
3.
Rotor Drive Hole
4.
Drive Pins
(Angled pins shown; pins can also be vertical or horizontal.)
5.
Drive Spindle Assembly
B
EC
KMAN
B
EC
KMAN
1
2
3
4
5
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......