Preinstallation Requirements
Preinstallation Requirements
A
4
Position the centrifuge so that the air diverter, shown in
, touches the wall behind the
centrifuge.
a.
Place the power cord to one side of the air diverter.
b.
To avoid damaging the power cord when installing or moving the centrifuge, be sure to
move the cord out of the way before pushing the centrifuge towards the wall.
WARNING
Do not place the centrifuge near areas containing flammable reagents or
combustible fluids. Vapors from these materials could enter the instrument’s
air system and be ignited.
Securing the Centrifuge to the Floor
Avanti J-26S series centrifuges are certified to meet the requirements of the European CE mark. To
meet these requirements, the centrifuge must be secured to the floor using the anchoring hardware
shipped with the instrument. This will prevent the centrifuge from moving in the unlikely event of
a rotor mishap.
Complete instructions for installing the anchoring kit are packaged with the hardware, which is
shipped with the centrifuge. The instructions (publication B10122) include a full-size template to be
used as a guide for drilling holes in the floor. Refer to this document for additional installation
instructions.
NOTE
Beckman Coulter representatives are not equipped to drill holes in your floor. The holes must be drilled
before
your scheduled installation.
Bio-Safety Level 3 Installation
For laboratories with epoxy aggregate (resinous) floors, such as BSL-3 labs, a non-invasive
installation kit (PN 393316) is available. The kit which consists of an adhesive-backed mounting
plate, is CSA certified for use on epoxy aggregate floors only.
Non-Invasive Centrifuge Restraint System for Vinyl Floors
For laboratories with vinyl tile floors or vinyl seamless floors, a non-invasive installation kit
(PN A84005) is available. The kit which consists of an adhesive-backed mounting plate, is CSA
certified for use on vinyl floors only.
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......