PN B05529AE
3-1
CHAPTER 3
Troubleshooting
Introduction
This section lists possible malfunctions, together with probable causes and corrective actions
required. Maintenance procedures are contained in
CHAPTER 4,
Care and Maintenance
. For any
problems not covered here contact Beckman Coulter Field Service.
NOTE
It is your responsibility to decontaminate the centrifuge, as well as any rotors and accessories, before
requesting service by Beckman Coulter.
User Messages
CAUTION
If the message SEr appears on the display, do not press any keys while the
message is displayed. Turn the centrifuge power off (O) and back on (I) to clear
the message. This message indicates that you have inadvertently accessed the
service mode. Pressing any keys while in this mode could erase the centrifuge
memory and critically interfere with future operation.
If a problem occurs during operation, the rotor decelerates to a stop, an error code appears on the
SPEED
display, and the centrifuge emits a series of audible tones to alert the operator to the error
condition. Such problems may result from incorrect input or from an equipment malfunction. Refer
to
Table 3.1
to determine the nature of the problem and recommended actions. If you are unable to
correct the problem, call Beckman Coulter Field Service. To help diagnose and correct the problem,
gather as much information about the situation as you can:
•
Write down the error number that appears on the display.
•
Note the operating situation when the error occurred (rotor in use, speed, load type, etc.).
•
Note any unusual environmental and/or operating conditions (ambient temperature, voltage
fluctuations, etc.).
•
Add any other information that may be helpful.
NOTE
Instructions for disabling the audible tones are in
CHAPTER 2,
Operation
.
Summary of Contents for Allegra X-30 Series
Page 4: ...PN B05529AE iv Revision History ...
Page 8: ...PN B05529AE viii Safety Notice Chemical and Biological Safety ...
Page 14: ...xiv Tables ...
Page 18: ...PN B05529AE xviii Introduction CE Mark ...
Page 30: ...PN B05529AE 1 12 Description Available Rotors ...
Page 56: ...PN B05529AE 4 6 Care and Maintenance Supply List ...
Page 62: ...PN B05529AE Warranty 2 Beckman Coulter Inc Allegra X 30 Series Centrifuge Warranty ...
Page 63: ......