![Beckman Coulter Allegra 6 Series Instruction Manual Download Page 41](http://html1.mh-extra.com/html/beckman-coulter/allegra-6-series/allegra-6-series_instruction-manual_2719690041.webp)
5-3
Care and Maintenance
• Clean the centrifuge case and door by wiping with a cloth damp-
ened with Solution 555. Do not use acetone or other solvents.
• Use a soft brush or vacuum to clean the air intake vent filter when
dust accumulates.
Air Intake
Filters
• (Nonrefrigerated models only) Remove the air-intake filters and
inspect for dust and dirt. Wash them using a mild soap or detergent
such as Solution 555. Rinse well and squeeze excess water out,
then air-dry before reinserting.
DECONTAMINATION
If the centrifuge and/or accessories are contaminated with radioactive
or pathogenic solutions, perform appropriate decontamination proce-
dures. Refer to
Chemical Resistances
to be sure the decontamination
method will not damage any part of the centrifuge.
STERILIZATION AND DISINFECTION
Ethanol (70%)
1
may be used on the centrifuge surface. See
Chemical
Resistances
for more information regarding chemical resistance of
centrifuge and accessory materials.
While Beckman Coulter has tested these methods and found that they
do not damage the centrifuge, no guarantee of sterility or disinfection
is expressed or implied. When sterilization or disinfection is a
concern, consult your laboratory safety officer regarding proper
methods to use.
TUBE BREAKAGE
If a glass tube breaks, and all the glass is not contained in the bucket
or rotor, thoroughly clean the interior of the chamber bowl.
!
CAUTION
Be careful when examining or cleaning the
gasket or chamber, as sharp glass fragments
may be embedded in their surfaces.
1
Flammability hazard. Do not use in or near operating centrifuges.