Secondary (manual)
Lock Lever
4-5
Troubleshooting
Figure 4-1. Emergency Access
Action
Result
5. Turn the hex wrench to
the right (clockwise) until
you hear a “click” or the
door opens.
The latch releases and the door can be opened.
If the rotor is still spinning,
close the door and wait until it stops
before attempting to remove it
.
!
WARNING
Never try to slow or stop the rotor by hand.
6. Open the door and remove