BE-IM-9
Allegra
®
6 Series
and Spinchron
™
R Centrifuges
Instruction Manual
Page 1: ...BE IM 9 Allegra 6 Series and Spinchron R Centrifuges Instruction Manual...
Page 2: ...lter les documents joints Atenci n consulte los documentos adjuntos Attenzione consultare le informazioni allegate On power Ein Netzverbindung Marche mise sous tension Encendido Acceso sotto tensione...
Page 3: ...uires removal of any covers can expose parts which involve the risk of electric shock or personal injury Make sure that the power switch is off and the centrifuge is disconnected from the main power s...
Page 4: ...amples that are pathogenic toxic or radioactive Such materials should not be used in this centrifuge however unless all necessary safety precautions are taken Observe all cautionary information printe...
Page 5: ...BE IM 9 February 2008 Allegra 6 Series and Spinchron R Centrifuges Instruction Manual 2008 Beckman Coulter Inc...
Page 6: ......
Page 7: ...ge Function 1 1 Allegra 6 and Spinchron Models 1 1 Specifications 1 3 Available Rotors 1 4 Safety Features 1 5 Name Rating Plate 1 6 Chassis 1 6 Housing 1 6 Door 1 6 Rotor Chamber 1 6 Drive 1 7 Refrig...
Page 8: ...ent 5 4 Temperature Knob Adjustment 5 6 Storage and Transport 5 7 Storage 5 7 Returning a Centrifuge 5 7 Supply List 5 8 Replacement Parts 5 8 Supplies 5 8 Warranty Figure 1 1 Allegra 6 Centrifuges 1...
Page 9: ...R centrifuges their functions specifications opera tion and routine operator care and maintenance We recommend that you read this entire manual especially the SAFETY NOTICE and all safety related info...
Page 10: ...t specifically recommended in this manual and or the appropriate rotor manual is the sole responsibility of the user CONVENTIONS Certain symbols are used in this manual to call out safety related and...
Page 11: ...O INTERFERENCE This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of FCC Rules These limits are designed to provide reasonable protecti...
Page 12: ...ronic Equipment WEEE Directive of the European Union The presence of this marking on the product indicates A28219 AA Note On the instrument the triangle background is yellow rather than gray 1 The dev...
Page 13: ...of small volume samples in multiwell plates for concentrating tissue culture cells cloning and replicate studies in vitro cytotoxicity studies receptor binding and genetic engineering experimentation...
Page 14: ...Allegra 6 Allegra 6R or Spinchron R Allegra 6KR 1 2 Description Figure 1 1 Allegra 6 Centrifuges...
Page 15: ...m 21 5 in Depth overall 66 0 cm 26 0 in 66 cm 26 in 66 0 cm 26 0 in Depth at base 58 0 cm 23 0 in 58 4 cm 23 0 in 58 4 cm 23 0 in Height door closed 40 6 cm 16 0 in 39 4 cm 15 5 in 67 3 cm 26 5 in Hei...
Page 16: ...ets rmax 204 mm Carriers rmax 163 mm 3750 3250 3210 1924 4 750 mL 4 192 mL GS6 TB 003 PTS 2000 Horizontal 4 place Sector base 100 mm tube rmax 170 mm Sector base 75 mm tube rmax 160 mm Tube base 16 mm...
Page 17: ...d the manual lock is in the UNLOCK position Two independent monitoring systems prevent the door from opening if the rotor is spinning An imbalance detector monitors the rotor during the run causing au...
Page 18: ...ort for speed verification In nonrefrigerated models the door has two filtered air intake ports for rotor cooling An electromechanical door lock system together with a manual lock prevents run initiat...
Page 19: ...centrifuge This air flow provides cooling for the rotor DRIVE The direct drive motor has permanently lubricated sealed bearings A tie down screw or rotor knob secures the rotor to the drive shaft The...
Page 20: ...nel Figure 1 3 nonrefrigerated model Figure 1 4 refrigerated models comprises indicators control knobs and switches POWER This two position rocker switch labeled I ON and O OFF controls electrical pow...
Page 21: ...deceleration to avoid remixing or resuspension of the sediment TIME This variable setting knob may be set for timed runs or longer duration continuous runs Either type of run can be stopped at any ti...
Page 22: ...ing temperature between 10 and 40 C in 1 C increments Recommended operating temperature range is from 5 to 25 C A thermistor in the chamber regulates the compressor to maintain chamber temperature 2 C...
Page 23: ...fluids Vapors from these materials could enter the centrifuge s air system and be ignited Never bring any flammable substances within the 30 cm 1 ft area surrounding the centrifuge Maintain a clearanc...
Page 24: ...ned to minimize the possibility of injury or damage that could result from centrifuge movement in the event of a rotor mishap The anchoring kit for a benchtop centrifuge contains bolts to secure the c...
Page 25: ...39 4 21 5 54 6 3 0 7 6 3 0 7 6 23 0 58 4 26 0 66 0 37 0 94 0 15 5 39 4 30 0 76 2 3 0 7 6 3 0 7 6 23 0 58 4 26 0 66 0 37 0 94 0 26 5 67 3 21 5 54 6 3 0 7 6 3 0 7 6 23 0 58 4 26 0 66 0 49 5 125 7 2 3 I...
Page 26: ...ire non grounding type of multiple outlet receptacle strip To ensure safety the centrifuge should be wired to a remote emer gency switch preferably outside the room where the centrifuge is housed or a...
Page 27: ...nding the centrifuge During operation you should not come within the 7 6 cm 3 in clearance envelope except to adjust centrifuge controls if necessary Do not lean on the centrifuge or place items on th...
Page 28: ...observe proper safety precautions for aerosol containment Use only the appropriate rotors and adapters Do not run toxic pathogenic or radioactive materials in this centrifuge without taking appropria...
Page 29: ...angle rotors either before or after installing the rotor in the centrifuge 6 Load the rotor according to the rotor manual Always run the rotor with a balanced load fill all positions on a swinging bu...
Page 30: ...otor install the lid and turn the knob to the right clockwise until firmly fastened on the shaft 8 Close the door Push firmly down until you hear a clicking latching sound indicating that the door is...
Page 31: ...celeration to set speed occurs The digital RPM display indicates the rotor speed The TIME control knob shows the time remaining in the run The C display indicates the chamber temperature refrigerated...
Page 32: ...s However for operator safety do not open the door until the rotor has come to a complete stop 3 Press the DOOR switch to OPEN CAUTION If disassembly reveals evidence of leakage you should assume that...
Page 33: ...ermine the nature of the problem and recommended corrective actions Possible causes for each problem are listed in the probable order of occurrence Perform the recommended corrective action in sequenc...
Page 34: ...to restart the run normally NOTE It is possible to trip the imbalance detector switch during rotor installation or loading Performing these actions with the power off will avoid the problem If the im...
Page 35: ...recool rotor Excess moisture in chamber Wipe out moisture before each run Tear or break in chamber gasket Call Beckman Coulter Field Service Refrigeration or electronic problem Call Beckman Coulter Fi...
Page 36: ...sibility of contact with a spinning rotor Turn the power off and disconnect the centrifuge from the main power source before proceeding Never attempt to override the door interlock system while the ro...
Page 37: ...rench to the right clockwise until you hear a click or the door opens The latch releases and the door can be opened If the rotor is still spinning close the door and wait until it stops before attempt...
Page 38: ......
Page 39: ...d rotor accessories WARNING Any maintenance procedure or servicing of this equipment that requires removal of any covers can expose parts which involve the risk of electric shock or personal injury Ma...
Page 40: ...the proposed method will not damage the equipment CLEANING Frequent cleaning will ensure proper operation and prolong the life of the centrifuge Always clean up spills when they occur to prevent corr...
Page 41: ...econtamination method will not damage any part of the centrifuge STERILIZATION AND DISINFECTION Ethanol 70 1 may be used on the centrifuge surface See Chemical Resistances for more information regardi...
Page 42: ...tal strobe Action Result 1 Install a rotor Reflective Tape 2 Secure the rotor tie down nut onto the drive shaft 3 Place a strip of reflective tape on half of the tie down nut 4 Shut the centrifuge doo...
Page 43: ...rpm 3 Check that the door is shut and securely latched Move the secondary latch lever to LOCK 4 Turn the TIME knob to HOLD The centrifuge will start 5 When the rotor speed stabilizes adjust the knob...
Page 44: ...E 10 0 10 20 30 40 3 Turn the power back on and set the TEMPERATURE control knob to 40 C 4 Check that the door is completely closed and the manual latch lever is in LOCK position 5 Let the centrifuge...
Page 45: ...ion a Returned Goods Authorization form must be obtained from Beckman Coulter Inc Contact your local Beckman Coulter office to obtain the RGA form and for packaging and shipping instructions RGA To pr...
Page 46: ...is given below Refer to the rotor manual for materials and supplies needed for rotors NOTE Publications referenced in this manual can be obtained by calling Beckman Coulter at 1 800 742 2345 in the U...
Page 47: ...will repair or at its election replace such component or accessory What consti tutes either reasonable service and a reasonable period of time shall be determined solely by Beckman Coulter REPLACEMENT...
Page 48: ......
Page 49: ......
Page 50: ......
Page 51: ...Beckman Coulter Inc 250 S Kraemer Blvd Brea California 92821 Sales and Service 1 800 742 2345 Internet www beckmancoulter com 2009 Beckman Coulter Inc All rights reserved...
Page 52: ......