2-2
Preinstallation Requirements
Figure 2-1. Centrifuge Dimensions
ELECTRICAL REQUIREMENTS
208-V, 60-Hz centrifuge . . . . . . . . . . . . . 187–229 VAC, 16 A, 60 Hz
200-V, 50/60-Hz centrifuge . . . . . . . . .190–210 VAC, 16 A, 50/60 Hz
230-V, 50-Hz centrifuge . . . . . . . . . . . . . 207–253 VAC, 13 A, 50 Hz
To reduce the risk of electrical shock, this centrifuge uses a 1.8-m
(6-ft) three-wire electrical cord (attached to the power connector at
the rear of the centrifuge) and plug to connect the centrifuge to earth-
ground. The plug for use in North America is the NEMA 6-20P; a
NEMA 6-20R socket (see Figure 2-2) should be available. (Contact
Beckman Coulter field service for specific information regarding
local requirements.) To preserve this safety feature: