Section 3: Service Procedures
Momentary power outage will not terminate the run. A Power diagnostic will be given for a power
failure at power glitch or outage. When a Power diagnostic occurs, all program parameters will be
saved and the Power diagnostic will be displayed on the VFD.
Diag.
#
Description
Action taken
When issued
Display
P1
Hardware power failure
indication
Shutdown with brake
On power up/ reset after
any power glitch or outage
P-1 Power failure
occurred, see manual
Door Diagnostics
Diag
#
Description
Action taken
When issued
Display
L1
Both door strikers not in place
Shutdown, no brake
When running or stopping
only
L-1 Door Latch,
call Service
L2
Left striker is not in place
Shutdown, no brake
When running, stopping or
pressing ENTER START
L-2 Door Latch,
call service
L3
Right striker is not in place
Shutdown, no brake
When running, stopping or
pressing ENTER START
L-3 Door Latch,
call Service
L4
Both door latches are open
Shutdown, no brake
When running, stopping or
pressing ENTER START
L-4 Door Latch,
call Service
L5
Left door latch open
Shutdown, no brake
When running, stopping or
pressing ENTER START
L-5 Door Latch,
call Service
L6
Right door latch open
Shutdown, no brake
When running, stopping or
pressing ENTER START
L-6 Door Latch,
call Service
L7
Cannot open the door
No action
When running, stopping or
pressing ENTER START
L-7 Can’t Open Door,
see manual
Imbalance Diagnostics
The Allegra X-12/R, X-15R centrifuge has a rotor imbalance detector as used in the Allegra 25/64R.
In a situation where an imbalanced rotor is being used an Imbalance diagnostic will be generated and
the instrument will brake the run to a stop. The system reports imbalance diagnostic if the sensor
reading is positive and the door is closed.
Temp Diagnostics
Diag.
#
Description
Action taken
When issued
Display
I1
Imbalanced rotor.
Shutdown with brake
I-1 Imbalance, see manual
I2
Bad imbalance detector
Shutdown with brake
Issued when imbalance
A/D does not change
(Connector not attached)
I-2 Imbalance Error,
call Service
I3
Imbalance not calibrated
Shutdown with brake
Issued when Imbalance
calibration value for rotor
is set to 0. This is the default
value when not calibrated.
I-3 Imbalance Error,
Call Service
I4
Imbalance sensor not calibrated
Run will not start
Issued when enter/start
pressed
I4 -Imbalance Error,
call Service
I5
Imbalance flag failure
Run will not start
Issued when enter/start
pressed
I5 -Imbalance Error, call
Service
I6
Imbalance flag calibration error
Run will not start
Issued when enter/start
pressed
I6 -Imbalance Error, call
Service
Allegra X-12/R, X-15R Service Manual 07/07
3-17
Summary of Contents for ALLEGRA X-12/R
Page 1: ...Part Number 679060 Rev AA ALLEGRA X 12 R X 15R Service Manual...
Page 18: ...SECTION 6 ACCESSORIES MISCELLANEOUS NO MATERIAL AVAILABLE AT THIS TIME...
Page 19: ...SECTION 7 APPLICATIONS NO MATERIAL AVAILABLE AT THIS TIME...
Page 20: ...SECTION 8 SERVICE MEMOS ISSUED DURING TRAINING CLASS...
Page 25: ...Section 1 Introduction Allegra X 12 R X 15R Service Manual 07 04 1 5...
Page 44: ...Section 3 Service Procedures Allegra X 12 R X 15R Service Manual 07 07 3 11...
Page 45: ...Section 3 Service Procedures Allegra X 15R 3 12 Allegra X 12 R X 15R Service Manual 07 07...
Page 60: ......
Page 61: ......
Page 62: ......
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......