Section 3: Service Procedures
Warning:
Unplug the centrifuge or turn off and lock out the main wall circuit
breaker before removing an instrument panel.
Warning:
As long as the yellow LED on the Motor driver board is lit, the Bus
Voltage is still present. Wait until LED is off.
3.1.4
Door Latch 392330, 392329 Replacement X-12/R
**Materials needed: Cord, Dacron (978139), Adhesive, Loctite 416 (6000808).
A.
Open front panel as described in 3.1.1.
B.
Remove harness from latch assembly.
C.
Remove the emergency release string from the latch by cutting the cord knot located
in the plug in the base of the instrument.
D.
Remove three nuts that hold the latch assembly in place.
E.
Remove latch
F.
Tie cord onto latch pivot bracket and secure it with Loctite.
G.
Replace latch, refer to BCI drawing number 392300 for adjustment of the latch.
H.
Tie both latch cords together so that both cords are the same length.
I.
Reinstall the latch cords in the plug in the base of the instrument.
J.
Make sure that the emergency release is functioning.
Warning:
Unplug the centrifuge or turn off and lock out the main wall circuit
breaker before removing an instrument panel.
Warning:
As long as the yellow LED on the Motor driver board is lit, the Bus
Voltage is still present. Wait until LED is off.
3.1.5
Door Latch 392900 Replacement X-15R
**Materials needed: Tether (392998), Door release, Adhesive, Loctite 416 (6000808).
K.
Open front panel as described in 3.1.1.
L.
Remove harness from latch assembly.
M.
Remove the emergency release string from the latch by cutting the cord knot located
in the plug in the base of the instrument.
N.
Remove two nuts that hold the latch assembly in place.
O.
Remove latch
P.
Tie cord onto latch pivot bracket and secure it with Loctite.
Q.
When installing the new latch it needs to be moved all the way to the left before
closing the door and adjusting the gap of .09 between latch and striker(BCI drawing
number 392930).
R.
Tie both latch cords together so that both cords are the same length.
S.
Reinstall the latch cords in the plug in the base of the instrument.
T.
Make sure that the emergency release is functioning.
3-4
Allegra X-12/R, X-15R Service Manual 07/07
Summary of Contents for ALLEGRA X-12/R
Page 1: ...Part Number 679060 Rev AA ALLEGRA X 12 R X 15R Service Manual...
Page 18: ...SECTION 6 ACCESSORIES MISCELLANEOUS NO MATERIAL AVAILABLE AT THIS TIME...
Page 19: ...SECTION 7 APPLICATIONS NO MATERIAL AVAILABLE AT THIS TIME...
Page 20: ...SECTION 8 SERVICE MEMOS ISSUED DURING TRAINING CLASS...
Page 25: ...Section 1 Introduction Allegra X 12 R X 15R Service Manual 07 04 1 5...
Page 44: ...Section 3 Service Procedures Allegra X 12 R X 15R Service Manual 07 07 3 11...
Page 45: ...Section 3 Service Procedures Allegra X 15R 3 12 Allegra X 12 R X 15R Service Manual 07 07...
Page 60: ......
Page 61: ......
Page 62: ......
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......