Omnisense 7000S/8000S
15-9
Metatarsal V
Figure 15.11 L-Gauge for use with metatarsal V
Position the L-gauge on the foot so that the inside corner is
at the dot marked previously, the long part is parallel to the
bone - extending to the lower part of the lateral malelus, and
the short part is parallel to the toes. Mark two perpendicular
lines, coinciding with the inside shape of the L-gauge; the
lines meet at the marked dot. Ensure that the gauge is fixed
to the malelus, and that the gauge’s internal corner remains
aligned with the dot marked earlier. The two lines drawn
will be referred to as the “long mark” and the “short mark.”
Figure 15.12 illustrates the marking procedure.
Figure 15.12 Marking the scan region with the L-gauge
Measuring SOS at the Metatarsal V
5.
Adjust the patient’s position so that the Achilles ligament is
supported by the leg rest, as close to the operator as possible.
6.
Use the CR probe for measuring at the metatarsal V.
7.
Disinfect the probe / attach probe cap
The L-gauge is used for
marking of the Region of
Interest at the metatar-
sal. (The same gauge
that is used for the pha-
lanx).
Draw two perpendicular
lines inside the L-gauge.
The distal prominence of
the metatarsal V is the
meeting point of the two
lines.
Summary of Contents for Sunlight Omnisense 7000S
Page 4: ...iv ...
Page 12: ...xii ...
Page 18: ...xviii ...
Page 22: ...xxii ...
Page 32: ...Omnisense 7000S 8000S 1 10 Conventions Used in this Guide ...
Page 108: ...Omnisense 7000S 8000S 5 14 Changing Users and Shutdown ...
Page 140: ...7 22 User Guide CHAPTER 7 Patient Files ...
Page 162: ...Omnisense 7000S 8000S 8 22 Measuring SOS at the Radius ...
Page 202: ...10 28 User Guide CHAPTER 10 System Administration ...
Page 216: ...Omnisense 7000S 8000S 12 4 Cleaning and Disinfection ...
Page 264: ...Omnisense 7000S 8000S 16 26 System Upgrade ...