14
3.2 Troubleshooting
If the pump assembly fails to indicate a pressure increase after considerable pumping
action of the handle, the following should be examined:
• Check to assure that the connections between the pump, the hose end and the
attached instrument(s) are tight and repeat operating instructions items 6 and 7
• Check to assure that all the unused output connectors are properly plugged.
If a pressure increase still cannot be obtained, it is possible that one or more of the seals
in the Pressure Pump is leaking and needs to be replaced.
Do not continue pumping if the functionality of the pump is not normal. Locate the fault before you
continue using the pump.
3.3 Available spare parts
The available spare parts include:
• Fine adjustment valve, complete set, code 8003186
• Pressure vent valve, complete set including the vent valve steel ball, code 8003187
• Vent valve steel ball, code 0500383
• 1/4” BSP female / 3/8” male adapter, code 8003189
• Plug 1/4” BSP male, code 7150120
3.4 Disposal
Incorrect disposal can put the environment at risk. Dispose of instrument components
and packaging materials in an environmentally compatible way and in accordance with
the country-specific waste disposal regulations.