Description
Manufacturer
Lubricant
Krytox®
GPL206
BeaconMedæs
P/N
GRS01-007
No. 2 Phillips screwdriver
Local source
No. 2 Flat-blade screwdriver
Local source
Oxygen compatible leak
Sherlock or equivalent
detector solution
Cleaning and Lubrication
BeaconMedæs outlets are cleaned for oxygen or the
indicated service. Exposed surfaces of outlets may be
cleaned with a mild detergent solution or wiped with
a disinfectant used for patient room surfaces that is
compatible with acetal and nylon plastic, anodized
aluminum, and die cast zinc.
Lubricate elastomer seals sparingly with Krytox®
GPL206. Refer to Tools and Service Materials.
DO NOT USE OIL.
Inspection and Testing
Medical gas outlets in high-use areas should be
inspected at least every six months and more frequently
if outlets are used daily. All outlets should be inspected
at least once a year.
Follow National Fire Protection Association
Standard 99, 99C “Gas and Vacuum Systems,”
and Canadian Standards Association, Standard Z
305.1 “Nonfl ammable Medical Gas Piping Systems”
requirements and methods for periodic testing and
maintenance of medical gas and vacuum systems.