5-5
“Oil-Less” Scroll Laboratory Air
4107 9006 25.11
5.0 Maintenance
5.2.6 2 - 3 Hp Service Intervals
2
OR with Tip Seal Kits used to support all systems
KIT NUMBER
WHERE USED
DESCRIPTION
QTY
CONTENTS
2903 0139 00
All Systems
Grease Pump Kit
1
Grease gun
Adaptors
2892 6100 20
All Systems
Grease Cartridge
1
Grease cartridge
1630 0311 00
7.5 Hp Units
15 Hp Units
Extension Nozzle for grease
pump
1
1
Extension nozzle
Notes:
1. The Grease Pump Kit does not include the Grease Cartridge or the extension nozzle for 7.5 Hp and 15 Hp
units. These items must be ordered separately.
2. 2 - 3 Hp Service intervas: are 5,000 hours
OR
every two (2) years, whichever comes fi rst, for the orbiting
bearings compressor lubrication and 10,000 hours
OR
every four (4) years, whichever comes fi rst, for the
pin crank bearings.
3. In high ambient conditions, the bearings must be greased more frequently: for every 9°F increase above
86°F, the maintenance interval should be reduced by 30%.
5.2.7 2 - 3 Hp Only: 10,000 Hours or Every Four (4) Years Compressor Tip Seal Kits
1
KIT NUMBER
WHERE USED
DESCRIPTION
QTY
CONTENTS
4107 4000 54
2 - 3 Hp Units
Kit - Tip seal 2 - 3 Hp
1 per
unit
Each kit contains:
(2) Sets of element seals
(1) Dust seal
(1) Round tube backer
seal
(1) Outlet pipe
Notes:
1. Service interval is 5,000 hours or every two years, whichever comes fi rst.
2. An authorized BeaconMedæs Service Technician should perform compressor lubrication and tip seals
replacement. Please contact BeaconMedæs Customer Service Department at 1-800-463-3427 prior to the
maintenance interval to schedule the maintenance.
IMPORTANT NOTE:
Greasing of the bearings of the compressor element must be done
with special grease and grease gun sold by BeaconMedaes below, and
according to a specifi c procedure. Contact BeaconMedaes for service.