4.0 Trouble Shooting
Problem
Possible Causes
Solution
Higher than usual differential
pressure across the system or
lower than usual flow through
the system
Dirty filter element
Too many filters offline
Blockages in system
See maintenance schedule in section 5.2
See operational procedures in sections 3.2-
3.4 for the minimum required filters to be
offline
Inspect tubes for blockages
Reduced vacuum level
Leak(s) in system
Check monthly for blockages and leaks in
the system. If a leak is suspected, pressurize
the system to maximum 30 psi and check
for leaks.
4-1
Medical Vacuum Filter Modules
4107 9022 25.00