Page 4 of 26
screws.
Turn switch off and then remove power cord.
Mounting information: The stability of the display must be guaranteed
following mounting of the foot/holder. The immersion depth of the
mounting screws has to be 10 to 12 mm including a 3 mm VESA mounting
plate. (See also table "Mounting screws" on the following). All these
requirements are satisfied when using the original foot. All
requirements must be observed when using customer-specific mounting
solutions.
Notice for users: The plug panel closed by the cover, must not be opened
by users.
Servicing information: If housing components have to be removed for
servicing, this must not be carried out in the presence of patients,
the user, or other persons not involved with servicing.
The following applies to installations in the USA and Canada: Molded
power supply plugs must comply with the requirements for "Hospital
Grade Attachments" UL 498.
Caution
Failure to observe the warnings may result in substantial damage to
property.
Provide sufficient heat dissipation
Holes are provided at the rear of the housing. The display must be
placed or secured on a hard, level surface at least 10cm from the wall
and 15cm away from other devices. Several displays can be butt-mounted
horizontally and vertically.
The following must be observed when mounting (VESA connection):
Mounting screws
Number
4
Thread
M4
Strength
12
Immersion depth
Min. 10 mm; Max. 12 mm
Torque
Max. 3 Nm
Please see 4.1.4 for details.