![Beacon 19LCDC16S User Manual Download Page 26](http://html1.mh-extra.com/html/beacon/19lcdc16s/19lcdc16s_user-manual_2716336026.webp)
Page 24 of 26
8.7 Climatic conditions
Operation
Ambient temperature range
+5 -- +35℃
Temperature gradient
Max. 7℃/h , no condensation
Relative Humidity
15%-85%
Atmospheric pressure
70 – 106 kPa
Transport and storage (packed)
Ambient temperature range
-20 -- +60℃
Temperature gradient
Max. 10℃/h, no condensation
Relative Humidity
10%-90%
Atmospheric pressure
70 – 106 kPa
8.8 Mechanical requirements
Operation
Vibration
According to EN 60068-2-6
10 ... 58 Hz with ± 0.075 mm deflection
58 ... 500 Hz at 10 m/s²
Shock
According to EN 60068-2-27 (single shock)
150 m/s², 6 ms
No permanent shock allowed in operating conditions
Packed unit
According to 2M2 EN60721-3-2
8.9 Safety specifications
Safety standards
EN60601-1、IEC60601-1、ANSI/AAMI ES60601-1:2005&CSA C22.2
No.60601-1:2008
Approvals
cTUVus、CCC、CB (NCB Lab.)
Protection class
Protection class I
Degree of protection
to DIN 40050
IP X0