4. There will be a long beep following the accomplishment of measurement. The
air in the cuff will deflate quickly and the blood pressure reading and pulse
reading will show in the display. The measuring time and date will also display,
alternating on the screen.
5. Press the button ‘
O/I
’ to return to the standby mode. Please rest for at least 3
minutes for another measurement. If the device isn’t used for 3 minutes, the
device will return to standby mode automatically.
7.1 Automatic Inflation
There are 4 levels of given inflation pressure for this device: 190mmHg, 230mmHg,
270mmHg and 300mmHg. When 190mmHg is not enough or movement of arm
occurs, the device will automatically inflate to reasonable pressure level to
ensure a successful measurement. It is not a fault.
7.2 Rapid Deflation During Measurement
If you do not feel well during measurement or want to stop the measurement for
some reason, you can press the
O/I
button. The device will quickly release the
air in cuff and the device will be returned to standby mode.
Repeated measurements with intervals of 3 minutes are recommended, so you
can calculate the average to get more accurate measurement. Atherosclerosis
patients are required longer intervals (10-15 minutes) as elasticity of patients’
vessels decreased significantly in these diseases. 10-15 minute intervals is also
applicable for patients suffering from diabetes for a long time.
8. Function of Memory
8.1 Memory Recall
1. RX800 can store 90 sets of readings, and will automatically calculate the average
value of the latest 3 readings. When the memory is full (90 sets of readings are
stored), the oldest reading will be replaced by the new one. Memory will not
clear away even if power supply is removed.