5. Install a sample tube on the SIP to ensure that the outer sleeve has been properly installed.
If the sleeve hits the bottom of the tube, loosen the retainer slightly and push the sleeve up as far as it will
go. Tighten the retainer.
Changing the sample tube O-ring
Introduction
This topic describes how to replace the sample tube O-ring.
The sample tube O-ring, located within the retainer, forms a seal that allows the droplet containment vacuum
to function properly.
When to replace the O-ring
Replace the O-ring when droplets form at the end of the sample injection tube while the vacuum is operating.
Cytometer hardware might be contaminated with biohazardous material. Wear suitable protective
clothing, eyewear, and gloves whenever cleaning the cytometer or replacing parts.
Procedure
To change the O-ring:
1. Remove the outer sleeve from the sample injection tube by turning the retainer counter-clockwise.
2. Slide the outer sleeve from the retainer. See the following illustration.
2
1
Number
Components
1
O-ring
2
Retainer
3. Invert the outer droplet sleeve and allow the O-ring to fall onto the benchtop.
If the O-ring does not fall out initially, hold the O-ring with your free hand and slide the outer sleeve to
remove the O-ring.
4. Place the new O-ring into the retainer. Make sure the O-ring is seated properly in the bottom of the retainer.
5. Replace the outer sleeve in the retainer.
48
BD FACSymphony™ A1 Flow Cytometer User's Guide
Summary of Contents for FACSymphony A1 Flow
Page 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Page 6: ......
Page 10: ...This page intentionally left blank ...
Page 24: ...This page intentionally left blank ...
Page 36: ...This page intentionally left blank ...
Page 50: ...This page intentionally left blank ...
Page 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Page 88: ...This page intentionally left blank ...
Page 96: ...This page intentionally left blank ...