BD EleVation Instructions For Use Manual Download Page 6

6

NOTE: 

Ultrasound image guidance should be used to confirm BD EleVation™ Probe position relative 

to the target region to be sampled.

NOTE: 

Do not leave the BD EleVation™ Breast Biopsy System unattended with the BD EleVation™ 

Probe in the primed state.

NOTE: 

During the initial positioning of the BD EleVation™ Probe, it is possible to advance the 

inner needle and cutting cannula into the tissue using the available PRIME/PIERCE option. During 
subsequent samplings, the physician may wish to forego this step.

NOTE: 

The BD EleVation™ Probe may be primed for piercing while inside the patient if the needle 

needs to be advanced to dense tissues or lesions.

4. 

Acquire sample by pressing the SAMPLE button. The sampling process proceeds automatically 

a.

 

BD EleVation™ Driver creates a vacuum

b.

  Cutting cannula of the BD EleVation™ Probe retracts to expose the sample notch

c.

 

Tissue is drawn to the sample notch by vacuum

d.

  Cutting cannula advances to cut the tissue 

e.

 

Tissue is transported to the sample container by vacuum 

During this process, the SAMPLE button continually illuminates red and the PRIME/PIERCE button is 
deactivated. After the tissue sample is deposited into the sample container, the BD EleVation™ Breast 
Biopsy System resets to acquire another sample, and the SAMPLE and PRIME/PIERCE buttons will 
illuminate green. 

NOTE:

 Do not twist the sample container or remove the sample container while acquiring tissue 

samples. If the BD EleVation™ Breast Biopsy System detects an error related to the sample container 
inhibiting the vacuum seal, the device will not work. To return the sample container into the BD 
EleVation™ Driver, clear any tissue underneath the sample container or from the BD EleVation™ Probe 
where the sample container is seated, insert the sample container back into the BD EleVation™ Probe 
and twist clockwise to ensure the sample container is properly in place. 

NOTE: 

The BD EleVation™ Half-Notch Support Cannula may be used for thin breasts and/or 

superficial lesions. If utilized, the optional BD EleVation™ Half-Notch Support Cannula (Figure 7) 
should be removed from the package using aseptic technique and replace the support cannula 
included with the BD EleVation™ Probe. The BD EleVation™ Half-Notch Support Cannula will reduce 
the length of the sample notch from 2cm to 1cm. 

5. 

To acquire multiple samples, repeat Step 4 above.

NOTE:

 Collecting multiple tissue samples may help minimize the risk of a false negative biopsy.

6. 

Turn the sample container counterclockwise to remove from the driver. Remove the tissue 
samples from the sample container by hinging it open. If additional samples are required, 
replace the sample container into the device by turning clockwise to lock into place. Ensure 
the sample container is secure before acquiring additional samples.

 

Figure 8 

Figure 9

 

To replace sample container

 

To remove sample container

Figure 10

NOTE:

 Removal of the sample container during the sample sequence will result in loss of vacuum.

Figure 11

NOTE:

 Prior to removing the BD EleVation™ Probe from the breast, the support cannula may be 

detached and remain in the breast to retain a track to the biopsy site to place a tissue marker. 

NOTE:

 To remove the support cannula from the BD EleVation™ Probe, depress the coaxial arm and 

twist the coaxial hub.

Figure 12

7. 

After obtaining the final tissue sample, remove the BD EleVation™ Probe from the breast and 
care for the incision site as appropriate.

NOTE: 

If experiencing difficulty removing the BD EleVation™ Probe from the breast, first detach 

integrated support cannula. Then, holding the support cannula in place, remove BD EleVation™ 
Probe from breast.

NOTE: 

The sample container may be removed before or after the BD EleVation™ Probe is removed 

from the BD EleVation™ Driver. 

8. 

Remove the BD EleVation™ Probe from the BD EleVation™ Driver by pressing down on the 
locking tab, sliding the BD EleVation™ Probe cover completely forward and then pulling the 
BD EleVation™ Probe straight up from the BD EleVation™ Driver (Figure 13).

Figure 13

9. 

The BD EleVation™ Probe, support cannula, sample container, and any BD EleVation™ 
Accessory used in this procedure is for single-use only. After use, these products may be a 
potential biohazard. Handle and dispose of in accordance with accepted medical practice and 
local, state, and federal laws and regulations. Dispose of sharps in designated sharps disposal 
containers; these are single-use devices and therefore should be disposed appropriately, not 
recapped or resterilized. 

Dead Space

1cm

Open

Closed

Sample Container

Summary of Contents for EleVation

Page 1: ...BD EleVation Sistema de Biopsia Mam ria Instru es de utiliza o OH9DWLRQ Dansk BD EleVation Brystbiopsisystem Brugsanvisning Svenska BD EleVation br stbiopsisystem Bruksanvisning Suomi BD EleVation Ri...

Page 2: ......

Page 3: ...ation Probes Green EVSC Additional Sample Containers For use with all BD EleVation Probes N A EVCover Instrument Cover Indications for Use The BD EleVation Breast Biopsy System is indicated to obtain...

Page 4: ...psy System should only be used by a physician trained in its indicated use limitations and possible complications of percutaneous needle techniques 2 Do not attempt to remove the cover or modify the d...

Page 5: ...sterile barrier or foreign material NOTE The BD EleVation Driver automatically recognizes when a BD EleVation Probe has been properly inserted and the BD EleVation Driver will perform a calibration t...

Page 6: ...cluded with the BD EleVation Probe The BD EleVation Half Notch Support Cannula will reduce the length of the sample notch from 2cm to 1cm 5 To acquire multiple samples repeat Step 4 above NOTE Collect...

Page 7: ...the wireless charging stand will illuminate green 4 Place the BD EleVation Driver in the wireless charging stand as shown in Figure 14 The BATTERY indicator on the BD EleVation Driver will illuminate...

Page 8: ...Hz Output Voltage 5Vdc Output Current Max 2 4A Removing the AC adapter from wall power shall serve as isolation mains Do not position the AC adapter and charging stand such that it is difficult to rem...

Page 9: ...5 V1 P D 3 5 E1 P 80 to 800 MHz D 7 E1 P 800 MHz to 2 7GHz Where P is the max power in watts and D is the recommended separation distance in meters Field strengths from xed transmitters as determined...

Page 10: ...HANDLING OR USE OF THIS PRODUCT Some states countries do not allow an exclusion of implied warranties incidental or consequential damages You may be entitled to additional remedies under the laws of y...

Page 11: ...ation avec les sondes 12G BD EleVation Bleu EVH14 Canule de support demi encoche 14G Pour utilisation avec les sondes 14G BD EleVation Vert EV10S Stylet 10G Pour utilisation avec les sondes 10G BD Ele...

Page 12: ...Vation Apr s rest rilisation la st rilit de la sonde BD EleVation n est pas garantie du fait d un degr ind termin de contamination pyrog ne ou microbienne potentielle pouvant provoquer des complicatio...

Page 13: ...rer la sonde BD EleVation L indicateur BATTERY s allume en vert lorsque la charge est suffisante pour effectuer une biopsie Laisser charger le guide BD EleVation pendant la nuit apr s l avoir utilis...

Page 14: ...u des l sions 4 Pr lever l chantillon en appuyant sur le bouton SAMPLE Le d roulement du processus de pr l vement est automatique a Le guide BD EleVation cr e un vide b La canule de coupe de la sonde...

Page 15: ...eVation bien s cher l air libre avant de le placer dans la station de chargement sans l Hormis un nettoyage r gulier aucun entretien pr ventif n est requis Il est toutefois recommand de renvoyer le di...

Page 16: ...pannage sp ci ques Si une condition d alerte persiste apr s avoir corrig des probl mes connus et remplac la sonde BD EleVation contactez un technicien agr NOTE si l erreur du syst me de biopsie mammai...

Page 17: ...ions RF CISPR11 Groupe 1 Classe A Le syst me de biopsie mammaire BD EleVation utilise l nergie RF uniquement pour sa fonction interne Par cons quent ses missions RF sont tr s faibles et ne sont pas su...

Page 18: ...3 6893 3 6893 7 3786 100 11 6666 11 6666 23 3333 Distances de s paration recommand es entre les quipements de communication RF portables et mobiles et le syst me biopsie mammaire BD EleVation quipeme...

Page 19: ...Sonden Blau EVH14 14G Halbkerbenst tzkan le Zur Verwendung mit 14G BD EleVation Sonden Gr n EV10S 10G Mandrin Zur Verwendung mit 10G BD EleVation Sonden Wei nicht zutreffend EV12S 12G Mandrin Zur Verw...

Page 20: ...kann nach einer erneuten Sterilisation nicht gew hrleistet werden da das Ausma einer evtl zu infekti sen Komplikationen f hrenden potenziellen Pyrogen oder Mikrobenkontamination nicht zu ermessen ist...

Reviews: