COVIDAIR S/T Easy Quick Start Guide
Page
6
Quick
Start
COVIDAIR
S/T
Easy
English
1.1.docx
/
BCD
/
30.04.2020
2.3. Start and Stop ventilation
Start and stop via the Touchscreen:
Press the button
to start ventilation.
The ventilation starts with the parameters saved in the "Parameters" and "Alarms" tabs.
The message "Ventilation activated" is displayed and the main button flashes each
time the patient breathes.
Press the button
to stop the ventilation, a message then asks for
confirmation of the stop and the message "Ventilation stopped" is displayed.
Start via the main push button:
Press the main push button to start ventilation.
The ventilation starts with the parameters saved in the "Parameters" and "Alarms" tabs.
The message "Ventilation activated" is displayed and the main
button flashes each time the
patient breathes.
Press the main push button again to stop the ventilation, a message then asks for
confirmation of the stop. The message "Ventilation stopped" is displayed.
NOTE
:
The ventilation start function using the main button works even if the device screen has not yet
started. It is therefore possible to start ventilation quickly with the last parameters saved!