COVIDAIR S/T Easy Quick Start Guide
Page
15
15
8. Setting the lower limit for the " Respiratory Rate " (RR) alarm in cycles per minute.
The value of the lower limit for the " Respiratory Rate " (RR) is between 4 and 52 by 1
cycle/min. increments.
9. Setting the upper limit for the " Respiratory Rate " (RR) alarm in cycles per minute.
The value of the upper limit for the " Respiratory Rate " (RR) is between 40 and 80 by 1
cycle/min. increments.
10. Activation slider for the "Pressure" alarm (Pressure Paw) (see its meaning and use at
the end of the chapter).
11. Adjustment of the low limit for the "Pressure". alarm (Pressure Paw) in cmH2O.
Setting the low limit for the "Pressure" alarm (Pressure Paw) in cmH
2
O.
The value of the low limit for the "Pressure" alarm (Pressure Paw) in cmH
2
O. The value
of the low limit for the " Pressure " alarm is between 0 and 20 by 1 cmH
2
O increments.
12 Adjustment of the upper limit for the "Pressure". alarm (Pressure Paw) in cmH
2
O.
The value of the upper limit for the " Pressure alarm " is between 40 and 80 by 1
cmH
2
O increments.
13. Activation slider for the "Apnea Time" alarm (Apnea) (see its meaning and use at the
end of the chapter)
14. Setting the time limit for the "Apnea time" (Apnea) alarm in seconds. The countdown
begins after 3 missed respiratory cycles. The limit value for " Apnea time" is between 10
and 60 by 5s increments.
15. Activation slider for the "Allowable leak time" (Leak) alarm (see its meaning and use at
the end of the chapter)
16. Setting the time limit for the "Allowable leak time" (Leak) alarm in seconds. The count
begins at the start of the leak. The limit value for the "Allowable leakage time" is
between 5 and 60 by 5.s increments.
17. Setting the low limit for the "Heart rate" (Fc)
in beats per minute alarm, can only be
changed if the pulse oximeter is used.
18. The value of the low limit for the "Heart rate" (Fc)
is between 40 and 80 by 1 pulse per
minute increments.
Setting of the upper limit for the "Heart rate" (Fc) in pulses per minute alarm, which can
only be modified if tthe pulse oximeter is used.
The value of the upper limit for the "Heart rate" (Fc)
is between 50 and 200 in 1 pulse per
minute increments.
19. Activation slider for the "Oxygen saturation" (SpO2) alarm, which can only be modified if
the pulse oximeter is used. (see its meaning and use at the end of the chapter)
20. Setting the time limit for the "Oxygen saturation" (SpO2) alarm in percent, only
modifiable if the pulse oximeter is used.
The limit value for "Oxygen saturation" is between 80 and 99 in 1% increments.
For each setting instruction, the "-" button allows you to decrease the value and the "+"
button allows you to increase the value by the indicated steps.
For each setting instruction, the "-" button allows you to decrease the value while the
"+" key button allows you to increase the value by the indicated steps. It is also possible
to let your finger pressed in order to reach the settings faster.