7
RF MEASUREMENT
RANGE
CURRENT
0 to 1000 mA RMS
POWER
0 to 500 W
ACCURACY
CURRENT
± 2% FS
POWER
± 5% FS
FREQUENCY
0 to 10 MHz
FUSE
500 mA Fast-Blow FWH-1/2
OSCILLOSCOPE OUTPUT
OUTPUT
Isolated
Uncalibrated
CONNECTION
BNC (50
Ω)
VARIABLE LOAD
VALUES
50 Ω
100 Ω
200 Ω
300 Ω
400 Ω
500 Ω
TYPE
Non-Inductive
ACCURACY
± 2% ± 1
Ω
POWER RATING
(MAX)
50 Ω
50 W
100 Ω
100 W
200 Ω
200 W
300 Ω
300 W
400 Ω
400 W
500 Ω
500 W
DUTY CYCLE
10 seconds on, 20 seconds off
LEAKAGE LOAD
VALUE
200
Ω
TYPE
Non-Inductive
ACCURACY
± 2
% ± 1 Ω
POWER RATING
50 W
CONNECTIONS
4 mm Banana
SPECIFICATIONS
Summary of Contents for BC Biomedical ESU-2000A
Page 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2000A USER MANUAL...
Page 2: ......
Page 8: ...vi This Page Intentionally Left Blank...
Page 17: ...9 NOTES...
Page 18: ...10 NOTES...
Page 19: ......