BC Group International BC Biomedical ESU-2000A User Manual Download Page 12

 

 

 
 
 
 
 
DUT RF Power Measurements 

 

1. 

Consult the DUT 

manufacturer’s manual for load specifications. 

 

2. 

Remove all connections between the DUT and the ESU-2000A Analyzer.  

 

3. 

Install the Switching Module on the ESU-2000A front panel so that the w

ords “ESU 

Power  Testing”  are  right-side-up.  When  installed  correctly,  the  Active  Electrode 
connectors  should  be  exposed,  and  the  RF  Leakage  connectors  should  be 
inaccessible. 

 

4. 

Connect  a  single  electrode  REM  cable  from  the  ESU-2000A  terminal  labeled 
“Patient Plate” to the DUT’s Patient Plate or Return Electrode. 

 

5. 

Connect a yellow safety lead from the DUT

’s RF Active Output to the desired load 

terminal of the ESU-2000A. The Active Electrode input connectors are color coded, 
and each corresponds to a color-coded range on the analog meter.  

 

6. 

Set the DUT for proper output mode and power level, being careful that the power 
setting does not exceed the full-scale Power range selected on the ESU-2000A.  

 

7. 

Activate the DUT output.  The Analyzer will indicate both mA RMS and Power (W) 
on the appropriate color-coded wattage meter range.  

 

8. 

Deactivate the DUT output before disconnecting from ESU-2000A, or before moving 
the Active Electrode test lead to a different input range connector. 

 
 

DUT RF Leakage Current Measurements 
 

1. 

Remove all connections between the DUT and the ESU-2000A Analyzer. 

  
2. 

Install the Switching Module on the ESU-2000A front panel so that the words 

“RF 

Leakage Testing” are right-side-up. When installed correctly, the Active Electrode 
connectors  should  be  inaccessible,  and  the  RF  Leakage  connectors  should  be 
exposed. 

 
3. 

Connect a yellow safety lead from the ESU-

2000A terminal labeled “RF Leakage” to 

the DUT’s RF Active Output. DO NOT MEASURE ACTIVE ELECTRODE LEAKAGE 
ON NON-ISOLATED ESUs with the ESU-2000A. Damage may occur to the ESU-
2000A. 

OPERATING INSTRUCTIONS 

NOTICE 

– PERFORMING TESTS 

 

REFER  TO  DUT  MANUFACTURER’S  SERVICE  MANUAL  FOR 
TEST PROCEDURES AND MEASUREMENT LIMITS

.

 

Summary of Contents for BC Biomedical ESU-2000A

Page 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2000A USER MANUAL...

Page 2: ......

Page 3: ...SPECIFICATIONS 7 NOTES 9 BC BIOMEDICAL ESU 2000A TABLE OF CONTENTS CALIBRATION INTERVAL To ensure the accuracy of the ESU 2000A BC Group International Inc recommends that it be calibrated at least on...

Page 4: ...ing the DUT to the ESU 2000A A serious hazard may occur if the patient is connected when testing with the ESU 2000A Do not connect any leads from the patient directly to the ESU 2000A or DUT WARNING M...

Page 5: ...the ESU 2000A while connected to the DUT as a hazardous condition may exist CAUTION ENVIRONMENT Exposure to environmental conditions outside the specifications can adversely affect the performance of...

Page 6: ...UAL FOR TEST PROCEDURES AND MEASUREMENT LIMITS NOTICE ABBREVIATIONS A Ampere s C Celsius Degree s DUT Device Under Test FS Full Scale Hz Hertz kg kilogram s lbs pounds M Mega 106 MHz Megahertz m milli...

Page 7: ...MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES NOTICE CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL INC 3081 ELM POIN...

Page 8: ...vi This Page Intentionally Left Blank...

Page 9: ...t testing The patient return electrode resistance simulation is adjustable in one ohm increments from 0 to 999 ohms The following are highlights of some of the main features LARGE 4 INCH ANALOG DISPLA...

Page 10: ...2 STANDARD ACCESSORIES BC20 00121 ACCESSORY KIT INCLUDES TEST LEADS ALLIGATOR CLIPS AND REPLACEMENT FUSE OPTIONAL ACCESSORIES BC20 30108 SOFT CARRYING CASE...

Page 11: ...that are present OVERVIEW Patient Plate Return Electrode Monitor Test Terminals RF Leakage Input Terminal Oscilloscope Output Fast Blow Fuse 4 INCH ANALOG DISPLAY Potentiometer Ohms ESU Ground RF Leak...

Page 12: ...ing does not exceed the full scale Power range selected on the ESU 2000A 7 Activate the DUT output The Analyzer will indicate both mA RMS and Power W on the appropriate color coded wattage meter range...

Page 13: ...00A 10 Connect the single electrode REM cable to the DUT s Patient Plate or Return Electrode and the Leakage Input on the ESU 2000A 11 Set the DUT for proper output mode and power level 12 Activate th...

Page 14: ...ENDED THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF SHIPMENT EXCLUSIONS THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WAR...

Page 15: ...PE OUTPUT OUTPUT Isolated Uncalibrated CONNECTION BNC 50 VARIABLE LOAD VALUES 50 100 200 300 400 500 TYPE Non Inductive ACCURACY 2 1 POWER RATING MAX 50 50 W 100 100 W 200 200 W 300 300 W 400 400 W 50...

Page 16: ...NVIRONMENTAL DISPLAY 4 1 2 Analog Meter Color coded display W and mA CONSTRUCTION ENCLOSURE ABS Plastic FACE PLATE Back printed Lexan SIZE 5 36 x 8 68 x 9 10 Inches 136 1 x 220 5 x 231 1 mm WEIGHT 4 L...

Page 17: ...9 NOTES...

Page 18: ...10 NOTES...

Page 19: ......

Page 20: ...NTERNATIONAL INC 3081 ELM POINT INDUSTRIAL DRIVE ST CHARLES MO 63301 USA 1 800 242 8428 1 314 638 3800 www bcgroupintl com sales bcgroupintl com ESU 2000A User Manual 5 20 Rev 09 Copyright 2020 Made i...

Reviews: