13
CODE
DISPLAY
FUNCTION
DESCRIPTION
BLOOD PRESSURE
62
0 mmHg
Zero Pressure
Puts all BP Channels at 0 mmHg
69
PA WEDGE
PA Wedge
Pulmonary Artery Wedge 10/2 mmHg
70 SWANGANZ Swan-Ganz
Insertion
Right Atrium CVP 15/10 Insert
Right Ventricular 25/0 Insert/Pull Back
Pulmonary Artery 25/10 Insert/Pull Back
Pulmonary –Arterial Wedge 10/2 Deflate
RESPIRATION
71 APNEA
Apnea
No
Respiration
72
RESP @ 10
Resp 10 BPM
Respiration at 10 BPM
73
RESP @ 20
Resp 20 BPM
Respiration at 20 BPM
74
RESP @ 40
Resp 40 BPM
Respiration at 40 BPM
75
RESP @ 80
Resp 80 BPM
Respiration at 80 BPM
76
RESP CVA
CVA Coincidence
NSR at 80 BPM, Respiration at 80 BPM
CALIBRATION
78
ECG RATE
Heart Rate Calibration
30, 40, 45, 60, 80*, 90, 100, 120, 140, 160, 180, 200,
220, 240, 260, 280, 300 BPM
81
2.5 Hz
Linearity/Speed
2.5 Hz triangle waveform
82
ECG SIZE
ECG Size Calibration
0.05, 0.10, 0.15, 0.20, 0.25, 0.30, 0.35, 0.40, 0.45, 0.50,
1.0*, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0, 5.5 mV
83
RR
Respiration Rate
0, 15, 20*, 30, 40, 60, 80, 100, 120 BrPM
84
RA
Respiration Amplitude
0.2, 0.50, 1.0*, 3.0 Ohms
85
BP1
BP1 Output Selection
ART, LV, RV, RA, -10, 0*, 80, 160, 240, 320, 400 mmHg
86
BP2
BP2 Output Selection
ART, LV, RV, LA, PA, RA, CVP, PAW, -10, 0*, 50, 100,
150, 200, 240 mmHg
87
BP3
BP3 Output Selection
ART, LV, RV, LA, PA, RA, CVP, PAW, -5, 0*, 20, 40, 60,
80, 100 mmHg
88
BP4
BP4 Output Selection
RV, PA, CVP, PAW, -5, 0*, 20, 40, 60, 80, 100 mmHg
89
ST
ST Elevation Selection
-0.8, -0.7, -0.6, -0.5, -0.4, -0.3, -0.2, -0.1, -.05, 0.0*, .05,
0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8 mV