PRE-DEFINED FUNCTION CODES:
A fixed function can be loaded by entering the
desired code. The key is pressed and then the
Number Keys are used to enter the code. The available
codes are located on the label on the back of the remote.
To make adjustments to the output selections,
and the Number Keys can be used to enter the
code to select output settings. The keys
are used to modify to the desired output. Any changes
made on the remote will be immediately made on the
Patient Simulator. The key returns the parameter
to its Standard settings.
CODE
DISPLAY
FUNCTION
DESCRIPTION
ATRIAL
15
AT TACH
Atrial Tachycardia
NSR of 160 BPM
18
PED TACH
Pediatric Tachycardia
Pediatric NSR of 160 BPM
BLOCK
19
1st AVB
1st Degree AV Block
80 BPM with a long P-R interval of 250 ms
20
RBBB
Right Bundle Branch Block
80 BPM with Normal P-wave and P-R interval but wider
QRS complexes
VENTRICULAR
27
BIGEMINY
Bigeminy
NSR of 80 BPM with every other beat a Type 1 PVC
28
U-PVC's
Unifocal PVCs
NSR of 80 BPM with 24Type 1 PVC’s per minute
29 M-PVC's
Multifocal
PVCs
NSR of 80 BPM with Type 1 and Type 2 PVCs (PVC
Type 1, 3 NSR, PVC Type 2, 3 NSR)