PROGRAMMABLE
PATIENT SIMULATOR
REMOTE CONTROL
PSR-2200-DS
USER MANUAL
Page 1: ...PROGRAMMABLE PATIENT SIMULATOR REMOTE CONTROL PSR 2200 DS USER MANUAL...
Page 2: ......
Page 3: ...1 DESCRIPTION 3 OVERVIEW 5 KEYS 6 SCREENS 7 MESSAGES 8 OPERATION 9 MANUAL REVISIONS 14 WARRANTY 14 SPECIFICATIONS 15 NOTES 16 BC BIOMEDICAL PSR 2200 DS TABLE OF CONTENTS...
Page 4: ...E IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILI...
Page 5: ...ATIONS AND ADJUSTMENTS LARGE GRAPHICS DISPLAY WITH BACKLIGHT DISPLAY BACKLIGHT WITH ADJUSTABLE TIMER PROGRAMMABLE AUTO SHUTDOWN TO CONSERVE BATTERY LIFE BC GROUP PSR 2200 DS PATIENT SIMULATOR REMOTE C...
Page 6: ...f the PSR 2200 DS and gives descriptions of the elements that are present OVERVIEW 20 Light Touch Keys 25 Coiled Cable For connection to a PC or Patient Simulator LCD Graphical Display with Backlight...
Page 7: ...5 List of Codes Battery Compartment 1 9V Battery Model and Serial Number Label...
Page 8: ...returns parameters to their Standard settings These keys are used to select the values of adjustable parameters as accessed through the key and codes When not in the Special Function mode these keys c...
Page 9: ...t Simulator The Backlight Time Screen allows for the setting of the Backlight Off Timer The Backlight can be Off set to 1 20 seconds or On The keys are used to choose the desired setting After 3 secon...
Page 10: ...messages COMMUNICATION ERROR This message indicates that the simulator did not acknowledge a command that was sent to it If the cables are connected properly cycle power on the PSR 2200 DS and the Pat...
Page 11: ...or on the simulator PRE DEFINED FUNCTION KEYS A fixed function can be loaded by pressing the desired key To make adjustments to the output selections and the Number Keys can be used to enter the code...
Page 12: ...by a P wave without a QRS 07 2AVB2 2nd Degree AV Block II 80 BPM with Normal P wave and P R interval but wider QRS complexes 08 3AVB 3rd Degree AV Block 80 BPM with P wave rate of 80 BPM and QRS rate...
Page 13: ...on the Patient Simulator The key returns the parameter to its Standard settings CODE DISPLAY FUNCTION DESCRIPTION ATRIAL 15 AT TACH Atrial Tachycardia NSR of 160 BPM 18 PED TACH Pediatric Tachycardia...
Page 14: ...ST 0 8 0 7 0 6 0 5 0 4 0 3 0 2 0 1 05 0 0 05 0 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 mV CODE DISPLAY FUNCTION DESCRIPTION PACEMAKER 32 A PACE Atrial 80 BPM with Pacer Pulse at the start of each P wave 33 AV P...
Page 15: ...n at 80 BPM CALIBRATION 78 ECG RATE Heart Rate Calibration 30 40 45 60 80 90 100 120 140 160 180 200 220 240 260 280 300 BPM 81 2 5 Hz Linearity Speed 2 5 Hz triangle waveform 82 ECG SIZE ECG Size Cal...
Page 16: ...ATE OF SHIPMENT EXCLUSIONS THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE...
Page 17: ...cklit ENCLOSURE 5 98 x 3 27 x 1 28 Inches 152 x 83 x 32 5 mm ABS Plastic WEIGHT 1 Lbs 0 45 Kg FACE PLATE Lexan Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE 20 to 65 C Electrical POWER Battery...
Page 18: ...16 NOTES...