7.14 Setting the Trigger
The beginning and end of volume calculations and respiratory
parameters are determined by means of a trigger. The trigger can
be set to start and stop according to flow or pressure in the high
flow channel (> 10 Measuring Respiratory Coefficients).
7.14.1 Selection of
Ventilation Mode
Over the selection of the ventilation mode you can adjust
reasonable standard trigger values for each mode. Using these
standard triggers 90% of all measuring tasks can be performed.
You can choose out of the following Ventilation modes:
Pediatric Ventilation (The flow measurement will be
taken on the Low Flow Channel where as the pressure
is measured on the connector for differential pressure
Pdiff)
Adult Ventilation (Measurements in the High Flow
Channel.)
High Frequency Ventilation (Measurements in the High
Flow Channel.)