17
Note 1 – The units for the pressure data is determined by the setting in Location 17. This may be set via the
Write command or manually using the Range Key. See Specifications Page for Ranges.
Note 2 – The units for the temperature data is determined by the setting in Location 18. This may be set via
the Write command or manually using the Range Key. See Specifications Page for Ranges.
The following is an example of how the Read/Write commands are used. For display
purposes, the symbol <cr> will be used to identify a carriage return ($0D), and the symbol
<lf> will be used to identify a line feed ($0A).
Data Sent
Data Returned
Meaning
R03<cr>
R03<cr><lf>
Echo of Command Sent
00010<cr><lf>
Auto Off is set to 10 minutes
W0300000<cr>
W0300000<cr><lf>
Echo of Command Sent
Change Auto off to disabled
LOCATION ACCESS
DESCRIPTION
RANGE
01
R
% BATTERY LIFE
REMAINING
0-100
02
R/W
CONTRAST
0-15
03
R/W
AUTO POWER OFF
0-30
04
R
RESERVED
0-65535
05
R
RESERVED
0-65535
06
R
RESERVED
0-65535
07
R
RESERVED
0-65535
08
R/W
FILTER CONSTANT
0-255
09 R RESERVED
0-65535
10 R
MODEL
0 = DPM-2001
1 = DPM-2001+
2 = DPM-2100
11
R
PRESSURE
See Note 1
12
R
MAX PRESSURE
See Note 1
13
R
MIN PRESSURE
See Note 1
14
R
TEMPERATURE
See Note 1
15
R
MAX TEMPERATURE
See Note 2
16
R
MIN TEMPERATURE
See Note 2
17 R/W
PRESSURE
UNITS
0=PSI
1=mmHg
2=inH2O
3=cmH20
18 R/W
TEMPERATURE
UNITS
0=Celsius
1=Fahrenheit