ENGLISH (US)
Page 21 of 28
OTH-SEM250-0356 Rev D
11.
Handheld Support
These data are included pursuant to IEC 60601-1 labelling requirements.
11.1.
Electromagnetic Environment
The Provizio SEM Scanners are intended for use in the electromagnetic environment specified in
Table 3. Electromagnetic Environment. Users should ensure that the Provizio SEM Scanners are
used in such an environment.
Table 3. Electromagnetic Environment
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF emissions
CISPR11
Group 1
The Provizio SEM Scanner 250 device uses RF
energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR11
Class B
The Provizio SEM Scanner 250 is suitable for use
in all establishments other than domestic and in
establishments connected to a low-voltage power
supply network which supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies