3.0 INSTALLATION

 

3.1 UNPACKING 

The scales come with-

 

Standard

 

KDAEC        KDAEC

 M       

KDAEC

 L 

√   

√   

√   

√   

√ 

KDAEC

 

Indicator with bracket and 2 screws 

The Platform 

Stainless Steel Pan 

Power supply Module 

An Instruction manual 

 

3.2 LOCATING 

 

The scales should not be placed in a location that will reduce 
the accuracy. 

 

Avoid extremes of temperature. Do not place the scale in direct 
sunlight or near air conditioning vents. 

 

Avoid unsuitable surfaces. The table or floor should be rigid 
and free from vibration. 

 

Avoid unstable power sources. Do not use the scale near 

large users of electricity such as welding equipment or large 
motors. 

 

Do not place the scale near vibrating machinery. 

 

Avoid high humidity that might cause condensation. Avoid 
direct contact with water. Do not spray or immerse the scales 
in water. 

 

Avoid air movement such as from fans or opening doors. Do 

not place the scale near open windows or air-conditioning 
vents. 

 

Keep the scales clean. Do not stack material on the scales 

when they are not in use.

 

 

To return to normal weighing, turn the scale off and back to on 

 

again or press the

 

 

key. 

13.2.1 FILTER

 

This parameter is for setting the speed of the display filter. For 

poor environments the filter should be set at its slowest to minimise 

external influences on the scale. For weighing small samples 

or gradual filling, the filter should be set at a faster setting. 

Press 

 

to scroll through the options. The display will show

 

FiL 1

” to “

FiL 3

”. If it is set to “

FiL 1

” then the display is at its 

slowest setting and at “

FiL 3

” the display is in its fastest setting. 

13.2.2 ZERO TRACKING 

This parameter is for setting the range of the zero tracking. Zero 

tracking will aid the scale to hold or return to zero and should be   

increased if large weights are left on the scale or temperature is   

not consistent. 

Press

 

 

to scroll through the settings. The display will show 

ZEo 1

” to “

ZEo 8

”. If it is set to “

ZEo 1

” the zero tracking is at its 

smallest range and “

ZEo 8

” the highest. 

13.2.3 STABILIZATION RANGE

 

This parameter is for setting the range of the stability indicator. This 

is used to determine when the scale will print automatically   

as well as indicate that the weight is stable. 

Press

 

 

to scroll through the settings. The display will show 

StA 1

” to “

StA 8

”. If it is set to “

StA 8

” then the stability is at its fastest 

and “

StA 1

” the slowest. 

16 

Summary of Contents for KDAEC 150

Page 1: ...KDAEC Instruction Manual KD AEC en V1 1 2011 ...

Page 2: ... Now the display will show the AD count press the key to enter in the factory calibration 5 The display show ZERO clear the other things on the platform and the press the key to confirm Place the correct calibration mass as selected by the user at the centre of the pan and press the key to confirm until the scale return to the normal weighting mode 19 TABLE OF CONTENTS 1 0 INTRODUCTION 1 2 0SPECIF...

Page 3: ...60 lb 16 oz x 5 oz Repeat ability 0 01 kg 0 02 lb 0 02 kg 0 05 lb 0 05 kg 0 1 lb 0 05 kg 0 1 lb 0 1 kg 0 2 lb Linearity 0 02 kg 0 04 lb 0 04 kg 0 1 lb 0 10 kg 0 2 lb 0 10 kg 0 2 lb 0 2 kg 0 4 lb 1 14 0 ERROR MESSAGES During the initial power on testing or during the operation the scale may show an error message The error messages are described below ERROR CODE DESCRIPTION POSSIBLE CAUSES nnnnn A c...

Page 4: ...section contains critical calibration reference information and is protected by a Pin Code which can only be accessed by a qualified technician To exit the user must switch off the scale when the display shows Pi 17 I Platform size KDAEC 30kg 75kg 150kg 390 x310 x 40mm KDAEC M 30kg 75kg 150kg 200kg 300kg 500 x500 x40mm KDAEC L 75kg 150kg 200kg 300kg 900 x600 x40mm POWER supply 4 x1 5VAA battery DC...

Page 5: ...gain or press the key 13 2 1 FILTER This parameter is for setting the speed of the display filter For poor environments the filter should be set at its slowest to minimise external influences on the scale For weighing small samples or gradual filling the filter should be set at a faster setting Press to scroll through the options The display will show FiL 1 to FiL 3 If it is set to FiL 1 then the ...

Page 6: ...ete list of parameters Switch off the scale Hold the key and then press momentarily Release the key The display shows the first technical parameter to set the filter FiL x To exit the parameter setting at any time press the key To scroll through the technical parameters press the key which will advance to the next parameter 15 3 3 SETTING UP NOTE When moving the scale take care to prevent bending ...

Page 7: ...ds the display for 15 1 x 15 seconds 3 Hti 2 Holds the display for 30 2 x 15 seconds 4 Hti 3 Holds the display for 45 3 x 15 seconds 5 Hti 4 Holds the display for 60 4 x 15 seconds Press the key to confirm the selection and move back tofirst parameter or press to exit the parameter setting HOLD AND PRINTING TABLE trn 1 trn 2 trn 3 Hod 1 USB communication is off Hold is off key has no function USB ...

Page 8: ...le if the hold function 13 9 0 AUTO POWER OFF The auto power off can be set by the user to disable the feature or to apre set time interval See setting of the parameter in section 13 1 1 10 0 OPERATIONS 10 1 SWITCHING ON THE SCALE To switch on the scale press the key The display will show the software revision number and then flash all the digits and symbols before counting down to zero This ensur...

Page 9: ... parameter 13 1 4 COMMUNICATION ADDRESS The fourth parameter is for setting the ID for the USB results output Display will show Add 0 DEFAULT SET This parameter sets the communication address which is sent via USB as an ID code There are 26 options to select from Add 0 to Add 25 Set Add 0 for no address The numbers relate to alphabets for example 1 A 2 B to 25 Y 13 1 5 SELECTION OF BAUD RATE The f...

Page 10: ...ter the scale returns to zero 13 1 3 ENABLING OF UNITS The third parameter is to enable or disable the weighing units so that the user can select the enabled units during the weighing operation Display will show on kg DEFAULT SET 1 on kg Enables the unit 2 off kg Disables the unit Press the key to confirm the selection and move to the next 11 Normal OutputU See section 13 1 7 add A G W 2 00 cr lf ...

Page 11: ...rror message CALEr will be displayed and the scale will return to zero Repeat the process correctly 9 Place the correct calibration mass as selected by the user at the centre of the pan Press the The display will return to weighing mode Note If the mass loaded is more than 20 of the factory calibration reference then an error message CALEr will be displayed and the scale will return to weighing wi...

Reviews: