Batch
code
Use by Date (last day of month)
Consult instructions for use
In Vitro
Diagnostic Medical Device
Technical Information
The C
ONTOUR
®
N
EXT
EZ blood glucose monitoring
system was tested by 115 lay users using capillary
blood samples and three C
ONTOUR
®
N
EXT
test strip
lots. The results were compared to the YSI
®
Glucose
Analyzer laboratory reference method, traceable to the
CDC hexokinase method. The tables below show how
well the two methods compared.
Table 1 – System accuracy results for glucose
concentration < 75 mg/dL
Difference range in values
between YSI laboratory
reference method and
C
ONTOUR
N
EXT
EZ meter
Within ±
5 mg/dL
Within ±
10 mg/dL
Within ±
15 mg/dL*
Number (and percent) of
samples within specified
range
6 of 7
(85.7%)
7 of 7
(100.0%)
7 of 7
(100.0%)
Table 2 – System accuracy results for glucose
concentration ≥ 75 mg/dL
Difference range in
values between YSI
®
laboratory reference
method and
C
ONTOUR
®
N
EXT
EZ
meter
Within ±
5%
Within ±
10%
Within ±
15%
Within ±
20%*
Number (and
percent) of samples
within specified
range
83 of 108
(76.9%)
104 of 108
(96.3%)
107 of 108
(99.1%)
108 of 108
(100.0%)
*Acceptance criteria in ISO 15197: 2003 are that 95% of all
differences in glucose values (i.e., between reference method
and meter) should be within 15 mg/dL for glucose values less
than 75 mg/dL, and within 20% for glucose values greater than
or equal to 75 mg/dL.