Bayer HealthCare Contour link User Manual Download Page 6

6

Important

Need Help ?  Bayer Diabetes Care: 1-800-348-8100

(available 24 hours a day, 7 days a week)

7

Important

www.bayerdiabetes.com/us

 Befor

e Y

ou Begin

  First Things First: Talking to Your Compatible 
Medtronic Device

Your 

C

ONTOUR

®

 L

INK

 meter is already set with the 

Send

 option 

turned on. (To turn the option off, see page 24.) Every time the 
meter displays a new blood glucose test result, the result is 
automatically sent to your compatible Medtronic device. 
However, your Medtronic device will only accept the glu-
cose result if the Meter Option in your device’s Utilities  
Menu is set to 

On

 and you have programmed the device with 

the meter’s ID (serial) number. Up to three different meter 
ID numbers, can be accepted. See your Medtronic device 
addendum that is included in your device box.

1.

   Program your meter ID number into your compatible 

Medtronic device. The ID number is the serial number 
found on the left side of the back label of your meter. It 
is a 6-digit numeric or alphanumeric number (0–9, A–F); 
for example, 120462 or B12B3C. 

See the User Guide 

for your Medtronic device for complete information on 
this step.

2.

   Set your meter to the same time and date as your Medtronic 

device (see page 25 in this User Guide).

The Medtronic device can receive a glucose result only if 
you have entered your meter’s ID number into the device. 
Conversely, no one else’s insulin pump can receive your 
glucose result, unless your meter ID number has been entered 
into that device. The meter uses a wireless signal to send the 
blood glucose result to the Medtronic device. However, the 

Send

 option must be set to ON (see page 24) and the meter 

and Medtronic device must be within 4 feet of each other. 
If you are not using the link functionality on your Medtronic 
device, you may want to turn the 

Send

 option off to save 

battery power.

Talking to Y

our Compatible Medtr

onic Device

Before You Begin

1.

   Have all the materials you will need ready before you 

begin  testing. This includes your 

C

ONTOUR

®

  L

INK

 meter, 

the 

C

ONTOUR

®

 

test strips, and the 

M

ICROLET

®

2 lancing 

device with 

M

ICROLET

®

 lancets. You may also need 

C

ONTOUR

 

control solution to run a quality control check.

2.

   Read your User Guide and all instructional materials 

provided in your meter kit before testing.

3.

   Let the meter and test strips adjust to the temperature 

of the location where you are testing. Whenever the 
meter is moved from one location to another, allow 20 
minutes for the meter to adjust to the temperature of 
that new location before performing a blood glucose 
test.

4.

   It is important not to use the test strips or control 

solution if the expiration date printed on the bottle 
label and carton has passed or if it has been six months 
(180 days) since you fi rst opened the control solution 
bottle. It  will  help to write the six month 

discard date

 

on the label in the area provided when you fi rst open 
the control solution.

5.

   

Before testing, always wash your hands with warm 
soapy water and dry them well.

WARNING: 

•  The meter and lancing device are for single patient 

use. Do not share them with anyone including other 
family members! Do not use on multiple patients!

1, 2

 

•  All parts of the kit are considered biohazardous and 

can potentially transmit infectious diseases, even 
after you have performed cleaning and disinfection.

For complete instructions on cleaning and disinfecting 
your meter and lancing device, please see page 16.

84378888_CntrLINK_UG_EN_FpBp_v3.indd   6-7

84378888_CntrLINK_UG_EN_FpBp_v3.indd   6-7

3/30/12   11:58 AM

3/30/12   11:58 AM

Summary of Contents for Contour link

Page 1: ...LINK WIRELESS BLOOD GLUCOSE MONITORING SYSTEM USER GUIDE USES BAYER S CONTOUR BLOOD GLUCOSE TEST STRIPS 93 ...

Page 2: ...ood Glucose Monitoring System is not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates Important Safety Instructions The MICROLET 2 lancing device included in this kit is not intended for use on multiple patients by health care professionals Neither the CONTOUR LINK meter nor the MICROLET 2 lancing device or lancets should be used with more th...

Page 3: ...rference and 2 This device must accept any interference received including interference that may cause undesired operation Warning Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment This portable transmitter with its antenna complies with FCC IC RF exposure limits for general population uncontrolled expos...

Page 4: ...p lid MICROLET 2 Lancing Device Use for fingertip puncture with Bayer s MICROLET lancet BATTERIES Use two 3 volt DL2032 or CR2032 lithium batter ies Replace when necessary or will appear if the meter is in 12 hour setup Displays the date format Results are displayed along with the preset units of measure The 7 appears when viewing the number of results over the past 7 days Average Indicates the ba...

Page 5: ...lator which calculates an estimated bolus to support your food intake or to correct a high blood glucose Your CONTOUR LINK meter can also send your fingertip blood glucose result directly to your compatible MiniMed Paradigm REAL Time Revel System eliminating the need to manually enter the result when calibrating the system Your Meter Features Important Things to Know Your CONTOUR LINK blood glucos...

Page 6: ... must be within 4 feet of each other If you are not using the link functionality on your Medtronic device you may want to turn the Send option off to save battery power Talking to Your Compatible Medtronic Device Before You Begin 1 Have all the materials you will need ready before you begin testing This includes your CONTOUR LINK meter the CONTOUR test strips and the MICROLET 2 lancing device with...

Page 7: ...id immediately after you have removed the test strip Excess moisture could damage the test strip Hold the test strip with the gray end facing up Insert the gray end into the test strip port on the meter No need to enter a code Check battery status If bat tery shows low see page 41 The meter will turn on A test strip with a flashing blood drop will appear letting you know the meter is ready to test...

Page 8: ...ip of the test strip to the drop of blood The blood is pulled into the test strip through the tip Do not press the tip against the skin or place the blood on top of the test strip Hold the tip of the test strip in the blood drop until the meter beeps After the beep you will see the meter count down the five seconds until the test is complete and your result is displayed Testing Your Blood WARNING ...

Page 9: ...all electronic devices on an aircraft 1 Insert a test strip into the meter to turn the meter on Before you begin your test press and hold the or key until is displayed The displayed next to the date will disappear indicating that the Send option is temporarily turned off 2 Test your blood as usual The result will not flash indicating that it is not being sent 3 When you turn the meter on to perfor...

Page 10: ...tion dosages health stress or activity Non diabetic expected values are 75 to 110 mg dL fasting 3 You should consult with your health care professional for glucose values specific to your needs Testing Your Blood Testing Your Blood Removing the used lancet In one hand hold the lancing device with your thumb on the grip indent with your other hand hold the gray endcap dial as shown and gently snap ...

Page 11: ...ice malfunctions for any cause or if you notice any changes in the meter case or display Cleaning and Disinfection Clorox Germicidal Wipes are available for purchase online at http www officedepot com or http www amazon com or for more information call Bayer Diabetes Care Customer Service at 1 800 348 8100 available 24 hours a day 7 days a week For more information see FDA Public Health Notificati...

Page 12: ... WET FOR 60 SECONDS Using a new Clorox Germicidal Wipe carefully wipe all outer surfaces of your meter until wet Do not allow cleaning solution to run into the meter through areas such as around the buttons or the meter s test strip or data ports Data Port Test Strip Port Buttons 2 Dry all meter surfaces and test strip port using a clean paper towel if needed Disinfecting Your Lancing Device Suppl...

Page 13: ...Excess moisture could damage the test strip NOTE Check the expiration and discard date Make sure the test strip does not appear torn or damaged Hold the test strip with the gray end facing up Insert the gray end into the test strip port on the meter Control Solution Testing Disinfecting Your Lancing Device cont d FOR PROPER DISINFECTION YOU MUST KEEP ALL ENDCAP SURFACES WET FOR 60 SECONDS 3 Using ...

Page 14: ...ur control test result is displayed The meter will automatically recognize and mark the control result for you NOTE The does not indicate the control test result is within the specified range If your control test result is out of range do not use your meter for blood glucose testing until you resolve the issue See What You Should Do on page 38 Compare your control test result with the Normal Contr...

Page 15: ...ions You can set your meter to display different options To get into the Setup mode press and hold M for three seconds A complete display will briefly ap pear then the Send option will be displayed NOTE You can exit the Setup mode at any time by pressing and holding M for 3 seconds All selections you have made up to that point will be saved Setting the Send Option The Send arrow will be displayed ...

Page 16: ...pear month day or day month Press M to set Setting the Sound Setting the Time Date and Sound Setting the Markers and Test Reminder Alarm You will now see three markers displayed on your meter along with the word Your CONTOUR LINK meter can give you the option of marking your test results by turning this option on See pages 28 33 to learn how to use these options Press or to turn the marker option ...

Page 17: ...strip in the meter until you have finished marking the results Press M or remove the used test strip to automatically save the test result in the meter memory OR Press or until all three markers disappear from the display Press M and remove the used test strip Dispose of the used test strip and lancet properly to prevent injury or contamination to others If you want to mark the test result If you ...

Page 18: ...ess any button to turn your meter off The meter will only set one alarm It will remember the last alarm set Any previous alarm setting will be lost Test Reminder Alarm To Cancel a Test Reminder Alarm Press M to enter the meter memory The 14 day average will appear along with the Press M again and a box will appear and flash around the Press or The empty box will now indicate the alarm has been tur...

Page 19: ...Alarm Press or ON Press M Alarm is set to go off after a set amount of time to remind you to run a post meal test Post Meal Press or ON Press M Result is marked as a post meal test Logbook Press or ON Press M Result is marked in memory as unique You should note this in your Logbook Using the Markers and Test Reminder Alarm Using the Markers and Test Reminder Alarm 84378888_CntrLINK_UG_EN_FpBp_v3 i...

Page 20: ...number of high results above 180 mg dL and low results below 72 mg dL during that seven day period Viewing Stored Results Press to review previous results Previous results Results will be shown starting with the most recent Each result will show the date and time the test was taken as well as any symbols you chose to mark the test result You can press or to scroll forward or backward through the r...

Page 21: ...the test strip and retest using a CONTOUR test strip If you do not resolve the problem contact Bayer Diabetes Care Customer Service at 1 800 348 8100 Error Codes and Symbols WHAT YOU SEE WHAT IT MEANS WHAT YOU SHOULD DO E 10 Invalid date or time See Setup instructions page 25 to set the date and time on your meter If you continue to have problems contact Bayer Diabetes Care Customer Service 1 800 ...

Page 22: ... result is still out of range retest with a new test strip and new control solution If result is still out of range call Bayer Diabetes Care Customer Service Some display segments do not appear when the meter initially turns on Possible meter electronics failure Compare meter display with the illustration on page 3 If the display is not correct contact Bayer Diabetes Care Customer Service Results ...

Page 23: ...y part numbers are DL2032 or CR2032 Caring for the System WARNING When a low battery warning displays on the meter replace the battery as soon as possible to avoid a test result not being stored in the meter memory Changing the Batteries Caring for the System When a low battery warning displays on the meter replace the battery as soon as possible to avoid a test result not being stored in the mete...

Page 24: ...atients to see how well the CONTOUR LINK compared to laboratory results For glucose results lower than 75 mg dL the percent and number of meter results that match the laboratory method within 15 mg dL 98 7 75 76 For glucose results at 75 mg dL or higher the percent and number of meter results that match the laboratory method within 20 98 5 319 324 Note When meter results are compared to the labora...

Page 25: ...s Alarm Beeps whenever a test strip is inserted into or removed from the meter when the test strip is filled with blood or control solution and when a test result appears in the display Two beeps will sound for an error Twenty beeps will sound when a programmed alarm goes off To assess intermediate measurement precision the performance of ten CONTOUR LINK meters was evaluated with three lots of se...

Page 26: ...om defects in materials and workmanship for 5 years from the date of original purchase except as noted below Dur ing the stated 5 year period Bayer HealthCare shall at no charge replace a unit found to be defective with an equiva lent or current version of the owner s model Limitations of Warranty This warranty is subject to the following exceptions and limitations 1 A 90 day warranty only will be...

Page 27: ...dicalDevices Safety AlertsandNotices ucm224025 htm 2 CDCClinicalReminder UseofFingerstickDevicesonMorethanOne Person Poses Risk for Transmitting Bloodborne Pathogens 2010 http www cdc gov injectionsafety Fingerstick DevicesBGM html 3 Longo DL et al Harrison s Principles of Internal Medicine 18th edition 2011 3003 Service Information If you have a problem and none of the problem solving steps in th...

Page 28: ...ICROLET Clinilog GLUCOFACTS and the No Coding logo are registered trademarks of Bayer No Coding is a trademark of Bayer Paradigm CareLink and Bolus Wizard are registered trademarks of Medtronic MiniMed Inc Revel is a trademark of Medtronic MiniMed Inc 84378888 2012 Bayer All rights reserved Rev 04 12 ...