9
6.2.2
Connection between the indicator and printer.
If there is one cable provided together with the printer, connect one end of the cable to the printer
interface and the other end to the 5-core metal socket of the indicator as following:
Indicator
Printer
1
3
2
7
TXD
S.GND
RXD
S.GND
Figure 8
* Note:
The printer will not print if the net weight is a negative value.
The print number varies from 0~99. When it exceeds 99, it will back to 0 automatically.
Insert power adaptor or turn on the battery switch and press to turn on the indicator. After the
indicator is turned on, it will first display and then the version number, then it will check the
strokes from “9” to “0” successively. If its connection is correct and zero position is normal, the indicator
will display . You should check the scale body if the indicator gives out a warning sound and
displays or , and then
-XXXXX
after. Under power-on state, press to turn it off. To
change the battery, see item 14.4 for details.
*Note:
The indicator will automatically enter basic weighing mode after power on.
7. POWER-ON OPERATION
I/O
POWER
You can convert between “basic weighing” and “check weighing” by pressing (MODE) key.
8. TWO WEIGHING MODES SWITCHING
MODE
SET
After power on, the scale will automatically enter basic weighing mode. Under this mode, the scale has
such basic functions as “ZERO”, “TARE”, “Weight unit converting”, “Arbitrary-point auto rectifying”,
“overload indicating” and so on.
9. BASIC WEIGHING MODE
10.1 Entering setup.
No matter whether it is in “basic weighing mode” or “check weighing mode”, you can make it to high and
low limit value setting process via operating keyboard. The method is as follows: Press (MODE) and
(
) simultaneously (you may also observe the former target value, high and low limit value and
percentage value in storage by using this method), at this time the indicator display XXXXX (XXXXX
refers to the previous target weight value in storage). If you do not need to amend it, press (ENTER) to
confirm, then go on to the next step. If need to revise, press (C) to make XXXXX turn to .
At this time, process sampling for target weight.
±
Δ
10. SETUP IN CHECK WEIGHING MODE
%
MODE
SET
ENTER
T
C
EXIT
BL
Target value
High limit value
Low limit value
Figure 9