FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
9-13
Calibration
Verifying Calibration Values
Verifying Calibration Values
Follow the instructions described in “Automatic Test Mode 1: Calibration
Mode 1” on page 9-14 and “Automatic Test Mode 2: Calibration Mode 2” on
page 9-16 for verification of pump calibration for isolation of failures or as
part of the annual preventive maintenance plan for the pump. Table 9-1
provides a list of the acceptable calibration ranges for each test mode.
Appendix B, “Calibration Verification Data Sheet” contains a blank data sheet
for recording the results from the calibration verification procedures
described below.
Table 9-1
Acceptable Calibration Ranges
Function
Acceptable Range
Calibration Verification (Verification in Automatic Test Mode 1)
Upstream Occlusion Calibration Value
3242-3314
Downstream Occlusion Calibration Value
2967-3039
Main Battery Calibration Value
688-714
Backup Battery Calibration Value
574-737
Calibration Verification (Verification in Automatic Test Mode 2)
NORM and MIN Air Sensor Calibration with Primed Tubing
NORM
330-565
MIN
NORM and MIN Air Sensor Calibration with Air Filled Tubing
NORM
< 11
MIN
NORM and MIN Air Sensor Calibration with No Tube
NORM
< 40
MIN
FSR Verification with Pump Door Open
RIGHT
500-820
LEFT
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...