9
Verifying Calibration Values
9-16
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
07-19-B1-729
Automatic Test Mode 2: Calibration Mode 2
The purpose of this Automatic Test Mode is to check the calibration of the air
sensor and to check the tube misloading sensor outputs.
1.
Open the door and check for a white tubing guide shim or the letter “A”
on the door opposite the air sensor.
Q
If the shim or “A” is present, use the air sensor calibration values of
330 - 565.
Q
If the shim or “A” is NOT present, perform tube guide installation per
“Tube Guide Shim Installation” on page 8-30.
2.
Plug the pump into a 115 ±5.0 VAC, 60 Hz power source. Enter
Automatic Test Mode 2:
2.1
Press and hold the
CLEAR TOT VOL
and
2
keys, then press the
ON-OFF/CHARGE
key.
2.2
Release the
ON-OFF/CHARGE
key without releasing the
CLEAR TOT VOL
and
2
keys for at least three seconds.
•
The pump is in Test Mode 2 if NORM and MIN are displayed
on the left hand side of the message (upper) display.
•
If the pump is not in Test Mode 2, turn off the pump and repeat
this step.
3.
Load a Baxter “s” suffix standard administration set and prime it with
fluid. Close the pump door. Ambient and solution temperatures must be
between 72° and 82° F (22° and 28° C).
4.
Open and close the pump door two more times and wait 2 minutes.
5.
Check the NORM value (at PRI RATE window) and the MIN value (at
PRI VTBI window).
6.
Remove the primed tubing and load an unprimed calibration tubing
segment and close the door.
7.
Open and close the door two more times and wait 2 minutes.
8.
View the NORM value displayed in the PRI RATE location at the upper
left of the pump’s main display.
9.
View the MIN value displayed in the PRI VTBI location at the lower left
of the main display. Both values must be less than 11.
Q
If they are within specification, proceed to the next step.
Q
If they are out of specification, perform the procedure in “Air Sensor
Calibration” on page 9-6.
10.
Remove the empty tubing segment and close the door.
Summary of Contents for Flo-Gard 6301
Page 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Page 251: ...07 19 B1 729 ...