Operator Manual for the Baxter ExactaMix 2400 Compounder
Page 100
53007410 Rev. C
6.
Check that:
•
Each syringe has fluid and its plunger is not stuck.
•
The appropriate inlet is used with each source container.
•
The inlets have no obstructions, kinks, tangles or plugs. If necessary, replace the
inlets. Refer to Changing the Tube Set on Page 42.
•
The occlusion detector is not damaged or dirty.
•
The outlet tube is straight and flat on the occlusion detector.
7.
At the
Fluid path will be flushed with UI
message, tap
OK
.
Message
8.
Check that the fluid moves properly during the flush.
9.
At the
UI flush complete
message, tap
OK
.
Message
10.
Remove the bag. Refer to Removing the Patient Bag on Page 90.
11.
Discard the bag.
Handling Other Interruptions and Errors
For more information about handling interruptions and errors, refer to Troubleshooting on
Page 193.