Operating Instructions
Level measurement
CleverLevel
®
LBFH/I
Point level detection – hygienic/industrial
Power supply
Output signal
Voltage supply range
8 ... 36 V DC
Output type
PNP
NPN
Digital (push-pull)
Reverse polarity protec-
tion
yes
Current load
100 mA max.
Current consumption
(without load)
25 mA type.,
40 mA max.
Short circuit protec-
tion
yes
Power-up
time
< 2 s
Voltage drop
PNP: (+Vs -0.5 V) ± 0.2 V,
Rload = 10 kΩ
NPN: (+0.5 V) ± 0.5 V,
Rload = 10 kΩ
Leakage current
± 100 μA max.
Switching logic
Normally open (NO), active
low
Normally closed (NC), high
enabled
www.baumer.com
13 / 48
Process conditions
Type
Process connection
BCID Process
temperature
continuous
[°C]
Process
pressure
[bar]
Process
temperature
t < 1 h
[°C]
Process
pressure
t < 1 h
[bar]
T
amb
< 50 °C
T
amb
< 50 °C
LBFI
G 1/2 A ISO 228-1
-1 ... 100
LBFI
1/2-14 NPT
-1 ... 100
LBFH/I
G1/2 A hygienic
G07
-40 ... 115
N02
-40 ... 115
A03
-40 ... 115
-1 ... 100 135
-1 ... 100 135
-1 … 10
135
-1 … 5
14. Configuration overview
14.1 Adaptive trigger
The adaptive trigger is a plug-and-play solution.
With the adaptive trigger, manual sensor
con
fi
guration is not required, since the trigger
will automatically con
fi
gure the output signal to
react to a new medium. The adaptive trigger is
very useful for applications with sticky media or
for applications where the medium is changed
frequently.
Normally, it is not necessary to con
fi
gure the
adaptive trigger, but if the advanced setup is
enabled, it is possible to de
fi
ne low and high
set points for the adaptive trigger window and a
damping. The steady detection function ensures
that the input signal is steady for 1 second before
the switch output reacts. If the steady detection
function is inactive, the switch output will react
immediately to each input change.
LBFI
G1/2 A hygienic
,
length 82 mm
-1
... 100
-1 ... 100
135
A03
-40 ... 115