the lens surface on both the X- and Y-screens. Perform dust check.
3.5 Replace Electrodes
Electrodes wear with use and also must be cleaned periodically due to
silica oxide buildup. It is recommended that the electrodes should be replaced
after 1,000 arc discharges. When the number of arc discharges reaches a count
of 1,000, a message prompting to replace the electrodes is displayed
immediately after turning on the power. Using the electrodes without a
replacement will result in greater splice loss and reduced splice strength.
Replace Electrodes
<1> Press
button on the operator panel until the LED indicator turn from
red to green, release the button
, now the machine is turned off.
<2> Remove the old electrodes and Method for tearing:
a. Loosen screw located on electrode cover.
b. Take electrode out of electrode cover.
(Electrode is fixed in the electrode cover)
<3> Clean the new electrodes with alcohol-impregnated clean gauze or lint-free
tissue and install them in the splicer.
<4> Place the electrode cover and tighten screw
●
Do not pull out night light wiring when replacing electrode.