4.2.2 

Programming mode

 

In the programming mode data is displayed to assist with parameter setting.   
 
When  SEL  and  MOV  keys  are  pressed  simultaneously  and  released  them 
immediately, the instrument can enter or exit of the mode. Press SEL key to 
select the content, they will be shown in turn. In each parameter, press MOV 
to choose the changeable digit,the chosen digit will become dark, and press 
INC to  change the value,  numbers  are shown 0-9 in  turns.  After setting the 
right values, press SEL to save and turn into next parameter. 
 
Please pay attention to the sign in the following examples. 
 

Significance 

Shown 

parameters 

Notes 

Weir type number 

(Must be set 

correctly) 

 

There are three types of weir can be chosen in this 

version BUFM660: Parshall flumes (ISO), Right-angle 

triangle weir, rectangular weir.   

This parameter displays in the form of “Pabc”, “P” is 
unchangeable, the latter three digits “abc” determines 

the chosen weir type number.   

Please referring to the follow “the weir type number 
table”. 

Installation height 

(Must be set 

correctly) 

 

The height is the distance from the bottom of 

transmitter to the bottom of the weir. 

The unit is M (meter). 

Full scale of 

instantaneous flow 

rate 

(Must be set 

correctly) 

 

The 20 mA output current corresponding instantaneous 

flow rate.   

Please set the value at first , the unit of the flow rate   

can be chosen in the latter parameters “Fuab”. 

This parameter displays in the form of five digits 

“abcd0”. The former four digits are changeable, the 
finial digit “0” is unchangeable. And the position of the 
decimal is decided by the first parameter “Weir type 
number”. 

Calibration of 

output current 

 

This parameter is used for correction of small output 

current error. The front two digits before the decimal 

point are used for correction of 20 mA, when the value 

increases, the 20mA output current will also increase. 

The latter two digits are used for correction of 4 mA, 

when the value increases, the 4 mA output will also 

increase, and vice versa.   

Summary of Contents for BUFM660 Series

Page 1: ...MANUAL Ultrasonic Open Channel Flow Meter...

Page 2: ...ing dimensions 9 3 Wiring 12 3 1 The host 12 3 2 Wiring diagram 12 3 3 The cable 13 3 4 The terminal blocks 13 4 Operation 14 4 1 Display and keys 14 4 2 Two working mode 14 5 Calibration 22 General c...

Page 3: ...wing functions High detection accuracy the flow meter measurement changes with 1mm the accuracy of change in level is 1 mm Suitable for a variety of weirs and flumes Parshall flumes ISO Right angle tr...

Page 4: ...3 1 3 Structure The BUFM660 series O C M consists of a probe and a host The remote control is optional...

Page 5: ...it can calculate the level as follows h H d Since speed of sound through air is affected by changes in temperature the BUFM660 O C M has integrated a temperature senor to improve accuracy For determi...

Page 6: ...0mA into 500 Ohms corresponding to instantaneous flow rate Relays outputs 6 multi function SPDT relays at most optional rated 5A 250VAC 30VDC high low and failsafe alarm and control corresponding to i...

Page 7: ...tallation considerations 2 1 1 hints for the host mounting The host should not be mounted in a confined space where temperatures may exceed the normal working temperature 40 70 if the host is mounted...

Page 8: ...he liquid to be measured But smooth vertical sidewalls weir tank will not cause false signals The probe must be mounted upstream of the flume or weir Do not over tighten the bolts on flange The stilli...

Page 9: ...8 Parshall flume triangular weir rectangular weir...

Page 10: ...9 2 2 Mounting dimensions The host...

Page 11: ...10...

Page 12: ...11 The probe The remote control...

Page 13: ...when the terminal cover is open 3 1 The host 3 2 Wiring diagram Remove the terminal cover to expose the terminals shown below The wiring instructions are on bottom of the terminals as below DC24V powe...

Page 14: ...elays can be provided with 5 pcs at most Carefully confirm the marked power supply terminals to ensure the correct power supply connection 3 3 The cable The cable between probe and host is 10m standar...

Page 15: ...ing some parameters 4 2 1 Running mode When the power is turned on the instrument takes several seconds to initialize and then runs into the running mode to start the normal level and flow measurement...

Page 16: ...and the unit for liquid level and space is M the unit for temperature is The output current is always corresponding to the instantaneous flow rate no matter which value disappears on the panel The max...

Page 17: ...keep flashing except the distance value the instantaneous flow rate and level value and the output current will not change The BUFM660 series can provide 6 relays which may be used for control purpos...

Page 18: ...e digits abc determines the chosen weir type number Please referring to the follow the weir type number table Installation height Must be set correctly The height is the distance from the bottom of tr...

Page 19: ...tatus of relay No 1 Correspond to instantaneous flow rate The format of this parameter is EC ab the front two digits EC are unchangeable the latter two ab determine the states of relay No 1 a correspo...

Page 20: ...cting On and off point value of relay No 4 Correspond to level in the weir Same with on and off point value of No 1 Correspond to level in the weir But when the relay No 4 is set as Accumulated Flow P...

Page 21: ...the setting and press SEL and MOV simultaneously to exit the setting mode the accumulated flow rate will be reset from zero Then enter the setting mode again and set this parameter as FC01 then save...

Page 22: ...Parshall flume 10 ft 1 07m 017 Parshall flume 12 ft 1 37m 018 Parshall flume 15 ft 1 67m 019 Parshall flume 20 ft 1 83m 1 504m 020 Parshall flume 25 ft 1 83m 1 314m 021 Parshall flume 30 ft 1 83m 1 17...

Page 23: ...tons firmly but not too hard to avoid damaging the circuit boards Also to avoid entering incorrect data do not push the buttons too fast When the instrument is used for pump control please avoid to di...

Page 24: ...the accumulate flow rate will increase too Adjust the full scale of instantaneous flow rate and test the output current with a multi meter the output current should change too Because the output curr...

Page 25: ...d pair cable and shielded cable is best When the baud rate is 9600bit s maximum length of the cable AWG26 specification and above is 1000m The RTU mode The RTU frame format the representation of data...

Page 26: ...data has 2 bytes and total 10 bytes That means the master requires 3 bytes instantaneous flow rate 1 byte the flow unit code 4 bytes accumulated flow rate and 2 bytes level value CRC Cyclic Redundanc...

Page 27: ...ue For example the slave response as follow 02H 03H 0AH 00H81H9DH 02H 00H00H0AH33H 01H8AH D5HF6H 02H 03H 0AH means the No 2 address instrument answer the response from the master with its order to sen...

Page 28: ...ontainer Add the water and the instrument will work well The instrument shows unstable or the measured value has a great deviation Wrong setting about the weir type number or the installation height o...

Page 29: ...28...

Reviews: