IFX-21

 

U

SER 

G

UIDE

 

I

NDUSTRIAL 

C

OMMUNICATIONS 

M

ODULE

 

 

 

 

 

 

 

 

 

Summary of Contents for EXACTUS IFX-21

Page 1: ...IFX 21 USER GUIDE INDUSTRIAL COMMUNICATIONS MODULE...

Page 2: ...t prior information and do not constitute the agreed contractual quality of the product It is the responsibility of the recipient of our products to ensure that any proprietary rights and existing law...

Page 3: ...ounting 6 Device Wiring 7 5 Device Operation and Setup 10 Overview of Connection Methods 10 Value Display and Indicating LEDs 11 Device Configuration 12 The Configuration Menu 13 Analog Output Operati...

Page 4: ...NTIFICATION The IFX 21 features a Nameplate as seen in Figure 1 FIGURE 1 EXAMPLE NAMEPLATE 1 Manufacturer and Product ID 2 Order Code and Serial Number 3 Power supply and current consumption 4 Ambient...

Page 5: ...l technical information about these ports 4 INSTALLATION AND COMMISSION SOFTWARE MINIMUM REQUIREMENTS Windows 7 8 1 and 10 compatible Windows 10 only for USB service port connection TemperaSure Versio...

Page 6: ...ir and or radiation shielding should be installed to ensure ambient temperatures remain within device operational limits see Section 7 1 In choosing a mounting location note should be taken that the d...

Page 7: ...available 4 4 2 TERMINAL BLOCKS The power analog output and Modbus terminals use detachable spring cage connector blocks To insert or release a wire push in the tab as shown in Figure 5 Pin out numbe...

Page 8: ...of 22 4 4 3 WIRING DIAGRAMS The IFX 21 enclosure features a simplified wiring diagram for all permanent device connections FIGURE 7 DEVICE WIRING Additional terminal wiring information is provided in...

Page 9: ...us TIA 485 A RS485 TABLE 1 EXPANDED DEVICE WIRING The IFX 21 is natively a passive analog device it will modulate incoming current but it provides no voltage If required the incoming power may be rout...

Page 10: ...ry to receive temperature readings over various protocols In these examples 4 units are shown however the methodology would be the same for any number of units BASF recommends that only communication...

Page 11: ...NG THE USB PORT VALUE DISPLAY AND INDICATING LEDS FIGURE 10 DEVICE DISPLAY 1 Process Temperature display 2 Error and power indicating LEDS During normal operation the IFX 21 will display the temperatu...

Page 12: ...nd adjust the pyrometer supply voltage Analog output settings and advanced operation modes such as peak picking or variance mode must be configured within the Pyrometer Please see the EXACTUS Optical...

Page 13: ...configured through the included TemperaSure software by selecting IFX Settings from the pyrometer channel menu 5 4 2 IFX 21 CONFIGURATION The IFX 21 has a configuration menu which allows the user to s...

Page 14: ...o Yes No 1 8mA Analog Output malfunction or failure No Yes No 2 2mA Default Output Yes Yes 2 7mA Analog Output is NOT enabled in Pyrometer settings Yes No Failure signal mA Default 3 6mA Yes Yes Measu...

Page 15: ...ere Not suitable for explosive or corrosive atmospheres TECHNICAL 7 2 1 1 ANALOG OUTPUT Absolute Accuracy 0 2 of reading 4 20mA Accuracy vs Ambient Temperature 15ppm C 4 20mA Connector Removable 4 pin...

Page 16: ...assigned Isolation 2 0kV Exposed Ethernet Ports 23 Telnet Port for configuration 502 Modbus TCP Port Max Link Distance 100 meters 328 feet Port Type Switch 7 2 1 3 PYROMETER PORT Connector RJ45 Elect...

Page 17: ...th pyrometer Isolation 700V Pin Assignments Pin 1 V Pin 2 V Pin 3 V Pin 4 V 7 2 1 5 MODBUS TIA 485 A RS485 Connector Removable 4 pin 5 00 5 08mm plug Electrical Protocol RS 485 Half Duplex Bit Protoco...

Page 18: ...INSTRUCTIONS Symptom Cause Remedy Display inactive No power supply connected Check power supply of the device Power supply applied device is defective Device must be replaced No analog output Analog o...

Page 19: ...device contains no user serviceable parts Devices returned to BASF must be shipped in protective packaging Use the Contacts listed in Section 1 for further instructions and to obtain an RMA number DIS...

Page 20: ...4 20 of 22 10 APPENDIX 7 SEGMENT DISPLAY 10 1 1 ALPHABET 10 1 2 MESSAGES INIT Shows on power up until pyrometer probe is discovered NO PROBE Shows when pyrometer probe fails to connect or is disconnec...

Page 21: ...S device has an 8 bit address All of the pyrometers are configured to respond to the MODBUS address 0x01 When using MODBUS TCP always leave the pyrometer address at 0x01 Use a unit ID UID of 1 in the...

Page 22: ...ated and no other MODBUS device on the RS485 network may be assigned either of those addresses If a network has multiple IFX21 devices they must be addressed at least 2 slave addresses apart i e first...