- 24 -
LO-BAT Indicator:
LO-BAT Indicator turns on when battery voltage is lower than normal operating voltage.
CAUTION!
When the scale is not used, power switch must be turned off.
5.4 Calibration
Calibration may be required when it is initially installed, if the scale is moved to a
substantial distance.
This is necessary because the weight of a mass in one location is not necessarily the
same in another location. Also, with time and use, mechanical deviations may occur.
1. Before entering calibration mode, first power is turned on, then press „
“ key
to turn the power on for 1minute.
2. Enter calibration mode
Press „
“ key to turn the power off. Press „
“ key first, then press „
“ key,
while keeping pressing „
“ key, the display will show the A/D value.
3. Calibration by the weight
Wait for the stable indicator „
“ to be displayed, press SET, the display will show
„00SAVE“, after 2 seconds, the display will return to A/D value.
NOTE!
SET key is located below the key „g/lb“. It is not marked on the key pad.
4. Place the full weight (15kg) on weighing platform. After the stable indicator
„
“ is displayed, then press SET, the display will show „00SAVE“ and now
calibration in finished.
5. Press „
“ key , turn off the power, press „
“ key turn on the power again,
place full weight (15kg) on the weighing platform to make sure whether
weighing is correct, if not, repeat steps 2-5.
6. Return to weighing mode.