BARTINGTON INSTRUMENTS
Page 15 of 19 OM2441/1
Note:
When using the 0.01Hz setting, allow sufficient time (up to 1 minute) for the filter to
settle. Momentary selection of the 1Hz setting will shorten the settling time.
Note:
The offset control can be used at the same time as using high pass filtering, but it
is not normally required. It is recommended that the offset control be switched to the OFF
position when using the HPF.
7.5. Using the SCU1 Outputs and Displays
7.5.1. Unconditioned and Conditioned Outputs
The SCU1 analogue outputs should be used when both the AC and DC content of the signals need
to be measured (e.g. with an oscilloscope, meter or an A-to-D system) or when a voltage output
is required for connection, for example, to a feedback or control system.
Note:
The
unconditioned
outputs should be used if access to the direct analogue output
signals from the magnetometer is required.
Note:
The
conditioned
outputs should be used to access the magnetometer output signals
after the application of filtering, offset and gain by the SCU1.
Consult the product brochure for maximum load.
To convert the analogue voltage level into units of magnetic field at the magnetometer, refer to
Converting Output Voltage to Magnetic Units
7.5.2. Display Panel Meters
The panel meters give a continuous digital display of the conditioned output voltage.