11-07-2013
P-800LT Freeze Point Analyzer
4-13
Programming
Abort — This selection immediately stops the analysis and prevents the Analyzer from proceeding. It is
intended as a troubleshooting tool only and should never be included as part of a normal Freeze point
measurement cycle.
Return — This should always be the last step in the State table. It tells the Analyzer to return to step 1.
Factory Default State Table
Step Time
Flush 90
Wait 5
Autozero
Cool 200
FP-ANAL 5
Return