Maintenance
75
03/2016
CFPP Process Analyzer CFPP-4.2
10.7
Checking the residual-current device (RCD)
Should only be performed by a qualified electrician.
Personal protective
equipmente
Safety goggles
Protective gloves (hazardous materials)
Required material
Software manual
Procedure
1.
Ensure that the ambient atmosphere is not explosive.
Warning!
Danger of explosion due to open ignition sources!
The control box contains ignition sources. Opening the control box can
cause explosion of potentially explosive atmosphere.
FOR THIS REASON:
•
Avoid measures see chapter 2 „Safety“.
1.
Turn the
Explosion protection
key switch to the
Off
position.
2.
Open the control box.
3.
Activate the test button
T
once every six month at least, if no other regional or
application-specific additional tests are specified.
The RCD must trigger.
10.7.1 Measures after the RCD has triggered
1.
If the RCD has triggered, switch it back on.
2.
Close the control box and secure the connections with safety bolts.
3.
Turn the
Explosion protection
key switch back to the
On
position.
Summary of Contents for CFPP-4.2
Page 1: ...CFPP Process Analyzer CFPP 4 2 Operating Instructions 395778MDHEN A 03 2016 ...
Page 6: ...IV Contents CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 7: ...General 5 03 2016 CFPP Process Analyzer CFPP 4 2 1 General ...
Page 9: ...Safety 7 03 2016 CFPP Process Analyzer CFPP 4 2 2 Safety ...
Page 25: ...Introduction 23 03 2016 CFPP Process Analyzer CFPP 4 2 3 Introduction ...
Page 28: ...26 Introduction CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 29: ...Technical Data 27 03 2016 CFPP Process Analyzer CFPP 4 2 4 Technical Data ...
Page 33: ...Analyzer Description 31 03 2016 CFPP Process Analyzer CFPP 4 2 5 Analyzer Description ...
Page 46: ...44 Transport Packaging and Storage CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 55: ...Initial Operation 53 03 2016 CFPP Process Analyzer CFPP 4 2 8 Initial Operation ...
Page 64: ...62 Initial Operation CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 65: ...Operation 63 03 2016 CFPP Process Analyzer CFPP 4 2 9 Operation ...
Page 69: ...Maintenance 67 03 2016 CFPP Process Analyzer CFPP 4 2 10 Maintenance ...
Page 102: ...100 Maintenance CFPP Process Analyzer CFPP 4 2 03 2016 ...
Page 103: ...Troubleshooting 101 03 2016 CFPP Process Analyzer CFPP 4 2 11 Troubleshooting ...
Page 107: ...Troubleshooting 105 03 2016 CFPP Process Analyzer CFPP 4 2 ...