34
Problem (cont.)
Possible Causes (cont.)
Action (cont.)
Display reads “Er5”.
PC board communication trouble.
Cycle power.
(Meter board data error)
Meter board not connected
Check connection from meter
to display/control board.
board (PCX70 10 pin connector)
to display/control board (J2).
Meter board (PCX70) out of
Replace meter board (PCX70).
tolerance.
Call Technical Service.
T
ROUBLESHOOTING