31
8 S
EGMENT
& 4
X
16 S
EGMENT
P
ROGRAMMABLE
M
ODELS W
/OTP
Steps 4 and 5 are used in the 4 program model only. If
you are using an 8 segment program, skip to step 6.
4.
Press the SCROLL button until display reads
“Hb.”
5.
Press the UP or DOWN button to toggle
between “bAnd, Hi, Lo and OFF.”
6.
Press the SCROLL button until display reads,
“tGt.”
7.
Press the UP or DOWN button to set a target
setpoint.
8.
Press the SCROLL button until display
reads,”rAtE.”
9.
Press the UP or DOWN button to select a
value in ramp units (seconds, minutes or
hours; set in the “rmP.U” parameter).
rmP.t
To ramp to a specified temperature at a set time:
1.
Press the PAGE button until you reach the
program list (ProG LiSt).
2.
Press the SCROLL button until display reads,
“tYPE.”
3.
Press the UP or DOWN button until display
reads, “rmP.t.”
4.
Press the SCROLL button until display reads,
“tGt.”
5.
Press the UP or DOWN button to set a target
setpoint.
6.
Press the SCROLL button until display reads,
“dur.”
7.
Press the UP or DOWN button to select a
time in ramp units (seconds, minutes or
hours; set in the “rmP.U” parameter.