INSTRUCTION MANUAL
DPS400
56-101-01180-C
1-4
Page 40
November 5, 2003
(c)
Slowly open the RATE CONTROL/LEAK TEST valve to set the static
pressure.
MANIFOLD PRESSURE
(INCHES OF MERCURY)
ALTITUDE
(FEET)
31 -985
30 -75
29 +860
28 1825
27 2815
26 3835
25 4890
20 10,730
15 17,905
10 27,375
MANIFOLD PRESSURE TEST
Table 3
(d)
When testing is complete, refer to Standard Test Set Shutdown Procedures to
return the test set to ambient pressures.