INSTRUCTION MANUAL
DPS400
56-101-01180-C
1-4
Page 16
November 5, 2003
(a) To begin setting the Pitot Port limits, press the MENU pushbutton. The Main
menu appears:
- Main Menu -
F1: Display Units
F2: Limits: Set, View, Enable/Disable
F3: Maintenance
F4: -Cancel-
F1
F2
LEAK
TEST
INC
F3
F4
CO NTRAST
DEC
MENU
PUMP
(b)
Press
F2
to set the pitot (or static) protection limits. The Limits menu appears:
L im its : V ie w , S e t, E n a b le / D is a b le
F 1 :
S ta tic L im its ( A lt, V S I, L e a k R a te )
F 2 :
P ito t L im its (A ir s p e e d , M a c h , L e a k R a te )
F 3 :
E n a b le /D is a b le
F 4 : -C a n c e l-
F 1
F 2
L E A K
T E S T
IN C
F 3
F 4
C O N T R A S T
D E C
M E N U
P U M P
Note:
Protection values are stored in non-volatile memory, and the values
displayed are those set in previous operations. It is recommended
that the operator always set their desired protection limits and
enable the protection circuits before each use.
(c)
Press
F2
to set the Pitot Port protection limits. The Pitot Limits menu appears:
Pitot Limits
F1: Airspeed (Including Zero Airspeed)
F2:
Mach
F3: Leak Rate Warning
F4: -Cancel
-
F1 F2
LEAK
TEST
INC
F3
F4
CONTRAST
DEC
MENU
PUMP
F1
sets the Airspeed protection limits.
F2
sets the Mach protection limits.
F3
sets the Leak Rate Warning.
F4