The StatLock® stabilization device may be included in PowerPICC® Provena™ Catheter kits. Please refer to Statlock® stabilization 

device Instructions For Use on the proper use and removal. The StatLock® stabilization device should be monitored daily and 

replaced at least every seven days.

The StatLock® Stabilization Device Procedure

Single Lumen

1. Secure catheter with 

StatLock® stabilization device.

2. Cover site and StatLock® 

stabilization device with 

transparent dressing.

3. Place anchor tape sticky side 

up, under hub. Wedge tape 

between hub and wings. 

4. Chevron anchor tape on top 

of transparent dressing.

Dual Lumen

1. Secure catheter with 

StatLock® stabilization device.

2. Cover site and StatLock® 

stabilization device with 

transparent dressing.

3. Place 1st anchor tape sticky 

side up, under one extension 

leg. Wedge tape between hub 

and wings. Chevron anchor 

tape on top of transparent 

dressing.

4. Place 2nd anchor tape sticky 

side up under remaining hub. 

Wedge tape between hub and 

wings. Chevron anchor tape on 

top of transparent dressing.

Tape Strip Stabilization Procedure

Single Lumen

1. Place 1st anchor tape over 

wings.

2. Cover site and 1st anchor 

tape with transparent dressing 

up to hub, but not over hub.

3. Place 2nd anchor tape sticky 

side up, under hub. Wedge tape 

between hub and wings. 

4. Chevron anchor tape on top 

of transparent dressing.

Dual Lumen

1. Place 1st anchor tape over 

wings.

2. Cover site and 1st anchor 

tape with transparent dressing 

up to hub, but not over hub.

3. Place 2nd anchor tape sticky 

side up, under one extension 

leg. Wedge tape between hub 

and wings. Chevron anchor 

tape on top of transparent 

dressing.

4. Place 3rd anchor tape sticky 

side up under remaining hub. 

Wedge tape between hub and 

wings. Chevron anchor tape on 

top of transparent dressing.

14. Verify Placement

A. 

PICCs should be positioned with the catheter tip in the lower 1/3 of the SVC.  Verify correct catheter tip position using 

radiography or other appropriate technology.

15. Power Injection Procedure

  WARNING:  PowerPICC® Provena™ Catheter indication for power injection of contrast media implies the catheter’s ability to 

with stand the procedure, but does not imply appropriateness of the procedure for a particular patient.  A suitably trained 

clinician is responsible for evaluating the health status of a patient as it pertains to a power injection procedure.

A. 

Remove the injection/needleless/end cap from the PowerPICC® Provena™ Catheter.

B. 

Attach a 10 mL or larger syringe filled with sterile saline.

Summary of Contents for PowerPICC Provena

Page 1: ...of their respective owners 2018 C R Bard Inc All rights reserved Manufacturer Bard Access Systems Inc 605 North 5600West Salt Lake City UT 84116 USA 801 522 5000 Clinical Information Hotline 1 800 443...

Page 2: ...ue injections Begin appropriate medical intervention immediately Do not wipe the catheter with acetone based solutions tincture of iodine or polyethylene glycol containing ointments These can damage t...

Page 3: ...ed with heparin flush solution As reported in literature anaphylactic or anaphylactic like reactions occur in a small percentage of the population during placement1 positioning1 flushing2 of central v...

Page 4: ...ng the following Air Embolism Bleeding Brachial Plexus Injury Cardiac Arrhythmia Cardiac Tamponade Catheter Erosion Through the Skin Catheter Embolism Catheter Occlusion Catheter Related Sepsis Endoca...

Page 5: ...rd C Insert the safety introducer needle into the desired vein and observe for flashback Alternate Technique A safety IV catheter may be used as an alternate to the safety introducer needle Remove the...

Page 6: ...out advance the catheter into the desired position Caution Never leave stylet or stiffening wire in place after catheter insertion injury may occur Remove stylet or stiffening wire and T lock as appl...

Page 7: ...chor tape with transparent dressing up to hub but not over hub 3 Place 2nd anchor tape sticky side up under hub Wedge tape between hub and wings 4 Chevron anchor tape on top of transparent dressing Du...

Page 8: ...their prevention or treatment Caution If CHG allergy is suspected confirmatory testing is recommended4 5 A Dressing Changes Exit Site Cleaning Caution Do not use scissors to remove dressing to minimi...

Page 9: ...s designed to aid in the placement of central venous catheters The stylet material provides internal reinforcement to aid in catheter placement In addition the Sherlock TLS stylet may be used with the...

Page 10: ...er insertion injury may occur Remove stylet or stiffening wire and T lock as applicable after insertion D For additional information refer to catheter and detector Instructions for Use for the inserti...

Reviews: